Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis

被引:211
作者
Bozkaya, D
Muftu, S
Muftu, A
机构
[1] Northeastern Univ, Dept Mech Engn, Boston, MA 02115 USA
[2] Tufts Univ, Sch Dent Med, Dept Restorat Dent, Boston, MA 02111 USA
关键词
D O I
10.1016/j.prosdent.2004.07.024
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. The external contour of an implant and the magnitude of occlusal loading can have significant effects on the load transfer characteristics and may result in different bone failure rates for different implant systems. Purpose. The goal of this study was to investigate the effects of external,geometry and occlusal load magnitude on bone failure modes for 5 commercially, available dental implant systems. Material and methods. Five different implant systems; Ankylos., Astra., Bicon, ITI, and Nobel Biocare. comparable in size, but different in thread profile and crest module shapes, were compared using the finite element method. Type II bone quality was approximated and complete osseous intergration was assumed. Occlusal loads of varying magnitudes (0 to 2000 NI) were applied On the abutments supporting single tooth restorations at 11.3 degrees from the vertical axis with a 1-mm offset. Total overloaded bone area, where tensile and compressive normal stresses fell Outside of the recommended limits of 100 and 170 MPa. respctively, investigated for different load levels. Results. For moderate levels of occlusal loads up to 300 N, the compact bone was not overloaded by any of the implant systems. At the extreme end of the occlusal load range (1000 N or more) the overloading characteristics of implants may be dependent on geometric shape. Conclusion. In general, overloading occurs near the superior region of compact bone,, in compression. and it is primarily caused by the normal and lateral components of the occlusal load. At the region of intersection of compact and trabecular bone, overloading occurs in tension due to the vertical component of the occlusal load. For excessive forces greater than 1000 N, the overloaded areas of the bone varied considerably among S different implants systems evaluated.
引用
收藏
页码:523 / 530
页数:8
相关论文
共 42 条
[1]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[2]  
ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
[3]  
Akça K, 2001, INT J ORAL MAX IMPL, V16, P722
[4]  
[Anonymous], 1999, CONTEMPORAY IMPLANT
[5]  
BARBER JR, 2002, ELASTICITY, P137
[6]  
Branemark P I, 1999, Clin Implant Dent Relat Res, V1, P2, DOI 10.1111/j.1708-8208.1999.tb00086.x
[7]  
Brunski J B, 1999, Adv Dent Res, V13, P99
[8]  
Brunski JB, 1997, IMPLANTS DENT ESSENT, P63
[9]  
Chow J, 2001, Clin Implant Dent Relat Res, V3, P166, DOI 10.1111/j.1708-8208.2001.tb00137.x
[10]   Evaluation of design parameters of osseointegrated dental implants using finite element analysis [J].
Chun, HJ ;
Cheong, SY ;
Han, JH ;
Heo, SJ ;
Chung, JP ;
Rhyu, IC ;
Choi, YC ;
Baik, HK ;
Ku, Y ;
Kim, MH .
JOURNAL OF ORAL REHABILITATION, 2002, 29 (06) :565-574