Trimethoprimium-phosphotungstate ion associate as ion-exchanger for trimethoprimium ion-selective electrode

被引:8
作者
Aboudan, M [1 ]
Shoukry, AF [1 ]
Issa, YM [1 ]
机构
[1] ALEPPO UNIV,FAC SCI,DEPT CHEM,ALEPPO,SYRIA
关键词
trimethoprim; PVC membrane electrode; potentiometric determination;
D O I
10.1080/00032719608000388
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A PVC membrane electrode selective for trimethoprimium cation based on trimethoprimium-phosphotungstate (TM)(3) PT ion associate is prepared. The electrode shows a linear response with Nernstian slope over the range of 2.7x10(-6)-10(-2)M. Calibration graph slope of the electrode is 59 mV/TM concentration decade when preconditioned by soaking in distilled water for 1 hour to 15 days. Tile electrode has working pH range of 1.8 to 6.3 and exhibits good selectivity for the trimethoprim with respect to a large number of inorganic cations and organic substances of biological importance. TM is determined successfully in pure solutions and in tablets using the standard additions methods.
引用
收藏
页码:19 / 28
页数:10
相关论文