Membranes for endotoxin removal from dialysate:: Considerations on feasibility of commercial ceramic membranes

被引:15
作者
Bender, H
Pflänzel, A
Saunders, N
Czermak, P
Catapano, G
Vienken, J
机构
[1] Univ Appl Sci Giessen, Inst Biochem Engn & Membrane Technol, Dept KMUB Biotechnol, D-35390 Giessen, Germany
[2] Univ Calabria, Dept Chem & Mat Engn, Arcavacata, Italy
[3] Fresenius Med Care, Bad Homburg, Germany
关键词
ceramic; dialysate; membranes; sterilization; water;
D O I
10.1046/j.1525-1594.2000.06514.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
As the quality of water in dialysis fluid varies considerably, dialysate is often contaminated by large amounts of bacteria and endotoxins. Membrane properties and operating pressures are acknowledged to give high-flux dialysis with bicarbonate the bacteriological potential to favor passage of endotoxin fragments from the dialysate into the blood stream. Therefore, a sterile dialysate will have to become a standard. Ultrafiltration across hydrophobic synthetic membranes was shown to remove endotoxins (and their fragments) from dialysis water by the combined effect of filtration and adsorption. However, each module can be used for a limited time only. Ceramic membranes may represent an alternative to polymeric membranes for endotoxin removal. In this article, we tested the capacity of different commercial ceramic membranes with nominal molecular weight cut-off down to 1,000 to retain endotoxins from Ps. aeruginosa. The tested membranes did not generally produce dialysate meeting the Association for the Advancement of Medical Instrumentation standard. When using aluminum-containing membranes, we detected aluminum leaking into the dialysate that could possibly be transported into the blood stream.
引用
收藏
页码:826 / 829
页数:4
相关论文
共 14 条
[1]
*ASS ADV MED INSTR, 1990, AAMI STAND, V3
[2]
BACTERIA-FREE AND ENDOTOXIN-FREE DIALYSIS FLUID FOR USE IN CHRONIC-HEMODIALYSIS [J].
BAMBAUER, R ;
WALTHER, J ;
MEYER, S ;
OST, S ;
SCHAUER, M ;
JUNG, WK ;
GOHL, H ;
VIENKEN, J .
ARTIFICIAL ORGANS, 1994, 18 (03) :188-192
[3]
BOMMER J, 1987, CLIN NEPHROL, V27, P278
[4]
Bonnie-Schorn E, 1998, WATER QUALITY HEMODI
[5]
Booth C, 1986, Dev Biol Stand, V64, P271
[6]
DIFELICE A, 1993, KIDNEY INT S, V41, pS201
[7]
EBBEN J, 1987, T AM SOC ART INT ORG, V33, P269
[8]
FILTRATION OF DIALYSATE USING AN ONLINE DIALYSATE FILTER [J].
FRINAK, S ;
POLASCHEGG, HD ;
LEVIN, NW ;
POHLOD, DJ ;
DUMLER, F ;
SARAVOLATZ, LD .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (11) :691-697
[9]
KLEIN E, 1990, ARTIF ORGANS, V14, P85
[10]
PERMEABILITY OF DIALYZER MEMBRANES TO TNF ALPHA-INDUCING SUBSTANCES DERIVED FROM WATER BACTERIA [J].
LONNEMANN, G ;
BEHME, TC ;
LENZNER, B ;
FLOEGE, J ;
SCHULZE, M ;
COLTON, CK ;
KOCH, KM ;
SHALDON, S .
KIDNEY INTERNATIONAL, 1992, 42 (01) :61-68