摘要
Secondary ion mass spectrometry (SIMS) can be confidently used to measure uranium isotopic ratios in single particles. Dense particles of known isotopic composition and size allow the precision and the accuracy of the applied procedure to be estimated. These particles can be obtained by dissolving standard reference uranium materials, nebulizing the solution in droplets of proper diameter and collecting the particles after the desolvation and calcination of the droplets. A new instrumental set up, based on a commercial vibrating orifice aerosol generator to generate monodisperse droplets of the solutions from four uranium oxide reference materials, is described. The droplets were dried and calcined in a sequence of three furnaces. The morphology of the monodisperse uranium oxide particles was studied by scanning electron microscopy. It was observed that the particles were nearly spherical and consisted of dense material. Their diameter distribution evidenced the presence of two populations mainly, the first showing a narrow distribution with a maximum centered at approximately 1 mu m. The first statistical moment ratios between the two populations remained practically constant at 1.24 +/- 0.01. This demonstrated that the second population was due to the formation of one particle from two droplets of solution (theoretical double mass = diameter ratio of 3 root 2 = 1.26). Secondary ion mass spectrometry was used to verify the isotopic composition of the produced particles. Typical accuracies of better than 0.4% for U-235/U-238 and a few percent for the minor isotopes have been achieved. For the determination of the U-236 content. the signal at mass M = 239 (due to (UH+)-U-238) was used to correct the (UH+)-U-235 contribution to U-236 at mass M= 236, greatly improving the accuracy of the 236/233 ratio with increasing enrichment of the U-235 isotope. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1565 / 1575
页数:11
相关论文
共 21 条
[1]   GENERATION OF MONODISPERSE AEROSOL STANDARDS [J].
BERGLUND, RN ;
LIU, BYH .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1973, 7 (02) :147-153
[2]   Use of secondary ion mass spectrometry in nuclear forensic analysis for the characterization of plutonium and highly enriched uranium particles [J].
Betti, M ;
Tamborini, G ;
Koch, L .
ANALYTICAL CHEMISTRY, 1999, 71 (14) :2616-2622
[3]  
BETTI M, 1997, P INT WORKSH STAT ME, P125
[4]  
CARSON PG, 1995, J AEROSOL SCI, V26, P535, DOI 10.1016/0021-8502(94)00133-J
[5]  
COOLEY JN, 1997, P 19 ANN ESARDA S SA, P31
[6]   ANALYTICAL-CHEMISTRY IN THE AFTERMATH OF THE GULF-WAR [J].
DONOHUE, DL ;
ZEISLER, R .
ANALYTICAL CHEMISTRY, 1993, 65 (07) :A359-&
[7]   Real-time analysis of individual atmospheric aerosol particles: Design and performance of a portable ATOFMS [J].
Gard, E ;
Mayer, JE ;
Morrical, BD ;
Dienes, T ;
Fergenson, DP ;
Prather, KA .
ANALYTICAL CHEMISTRY, 1997, 69 (20) :4083-4091
[8]   LASER-INDUCED MASS ANALYSIS OF SINGLE PARTICLES IN THE AIRBORNE STATE [J].
HINZ, KP ;
KAUFMANN, R ;
SPENGLER, B .
ANALYTICAL CHEMISTRY, 1994, 66 (13) :2071-2076
[9]  
KOCH L, 1999, 98 I TRANS EL, P26
[10]   LASER IONIZATION MASS-SPECTROSCOPY OF SINGLE AEROSOL-PARTICLES [J].
MURPHY, DM ;
THOMSON, DS .
AEROSOL SCIENCE AND TECHNOLOGY, 1995, 22 (03) :237-249