Subthreshold test pulses versus low energy shock delivery to estimate high energy lead impedance in implanted cardioverter defibrillator patients

被引:8
作者
Vollmann, D
Luethje, L
Zenker, D
Domhof, S
Unterberg, C
机构
[1] Univ Gottingen, Dept Cardiol & Pneumol, D-3400 Gottingen, Germany
[2] Univ Gottingen, Dept Cardiovasc Surg, D-3400 Gottingen, Germany
[3] Univ Gottingen, Dept Med Stat, D-3400 Gottingen, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 01期
关键词
implantable defibrillator; high voltage circuit; lead failure; impedance;
D O I
10.1046/j.1460-9592.2003.00071.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The high energy lead impedance is valuable for detecting lead failure in ICDs, but until recently shock delivery was necessary for high energy impedance measurement. This study compared the use of subthreshold test pulses and low energy test shocks to estimate the high energy impedance. Immediately after implantation of Ventak Prizm ICDs in 29 patients, the lead impedance was measured with five subthreshold (0.4 muJ) test pulses, 5 low energy (1.1 J) shocks, and two to three high energy (16 +/- 4.5 J) shocks. The mean impedances measured using high energy shocks, low energy shocks, and subthreshold pulses were 42.0 +/- 7.3 Omega, 46.5 +/- 8.1 Omega, and 42.4 +/- 7.1 Omega, respectively. The impedances measured using high and low energy shocks differed significantly (P < 0.0001), while those obtained by high energy shocks and low energy pulses did not (P = 0.63). According to-the Pearson correlation coefficient, the impedance measurements with subthreshold pulses and low energy shocks were both closely correlated (P < 0.0001) with impedance values determined with high energy shocks. However, while the impedance values tended to be higher when measured with low energy shocks, the concordance correlation coefficient (c) was higher for subthreshold test pulse versus high energy shock (c = 0.92) than for low versus high energy shock (c = 0.73). Furthermore, the intraindividual variability of impedance measurements was lower with subthreshold pulse measurements than with low energy shocks. Compared with low energy shocks, impedance measurement with subthreshold pulses has higher reproducibility and a higher correlation with the impedance obtained by high energy shock delivery. Safe and painless high energy impedance estimation with subthreshold pulses might, therefore, help to detect ICD lead failure during routine follow-up.
引用
收藏
页码:457 / 460
页数:4
相关论文
共 13 条
[1]   Defibrillator challenges for the new millennium: The marriage of device and patient - Making and maintaining a good match [J].
Friedman, PA ;
Glikson, M ;
Stanton, MS .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (06) :697-709
[2]   Routine arrhythmia inductions for ICD follow-up: Are they obsolete? [J].
Glikson, M ;
Friedman, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (06) :915-920
[3]   TROUBLESHOOTING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SYSTEM MALFUNCTIONS - THE ROLE OF IMPEDANCE MEASUREMENTS [J].
HADDAD, L ;
PADULA, LE ;
MOREAU, M ;
SCHOENFELD, MH .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (09) :1456-1461
[4]   UTILITY OF LOW-ENERGY TEST SHOCKS FOR ESTIMATION OF CARDIAC AND ELECTRODE IMPEDANCE WITH IMPLANTABLE DEFIBRILLATORS [J].
LEITCH, JW ;
YEE, R ;
KLEIN, GJ ;
JONES, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (04) :410-416
[5]   A CONCORDANCE CORRELATION-COEFFICIENT TO EVALUATE REPRODUCIBILITY [J].
LIN, LI .
BIOMETRICS, 1989, 45 (01) :255-268
[6]   Predictors and mode of detection of transvenous lead malfunction in implantable defibrillators [J].
Luria, D ;
Glikson, M ;
Brady, PA ;
Lexvold, NY ;
Rasmussen, MJ ;
Hodge, DO ;
Chugh, SS ;
Rea, RF ;
Hayes, DL ;
Hammill, SC ;
Friedman, PA .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 87 (07) :901-+
[7]  
McAnulty J, 1997, NEW ENGL J MED, V337, P1576
[8]   Implantable defibrillators impedance measurement using pacing pulses versus shock delivery with intact and modified high voltage lead system [J].
Mehdirad, AA ;
Stohr, EC ;
Love, CJ ;
Nelson, SD ;
Schaal, SF .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (03) :437-441
[9]   Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia [J].
Moss, AJ ;
Hall, WJ ;
Cannom, DS ;
Daubert, JP ;
Higgins, SL ;
Klein, H ;
Levine, JH ;
Saksena, S ;
Waldo, AL ;
Wilber, D ;
Brown, MW ;
Heo, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1996, 335 (26) :1933-1940
[10]   Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction [J].
Moss, AJ ;
Zareba, W ;
Hall, WJ ;
Klein, H ;
Wilber, DJ ;
Cannom, DS ;
Daubert, JP ;
Higgins, SL ;
Brown, MW ;
Andrews, ML .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (12) :877-883