The UBiT-100 13CO2 infrared analyzer:: Comparison between infrared spectrometric analysis and mass spectrometric analysis

被引:42
作者
Ohara, S
Kato, M
Asaka, M
Toyota, T
机构
[1] Tohoku Univ, Sch Med, Dept Internal Med 3, Aoba Ku, Sendai, Miyagi 98077, Japan
[2] Tohoku Univ, Sch Med, Otsuka UBiT Study Grp, Sendai, Miyagi 98077, Japan
[3] Hokkaido Univ, Sch Med, Dept Internal Med 3, Sapporo, Hokkaido 060, Japan
关键词
D O I
10.1046/j.1523-5378.1998.08046.x
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background. Although the C-13-urea breath test is the most accurate noninvasive method for detecting the presence of H. pylori infection, the requirement for an expensive mass spectrometer to analyze breath samples has prevented physicians from providing rapid results near the patient. Recently, a new type of infrared spectrometric analyzer, the UBiT-100, was developed for analyzing (CO2)-C-13 in breath. The purpose of this study is to compare results analyzed by the UBiT-100 to those analyzed by the mass spectrometric method. Methods. Four hundred and fifty-three subjects participated in this study. Breath samples were collected before administration of 100 mg of C-13-urea and at 10, 20, 30, 45 and 60 min after administration. Subjects were asked to hold their breath for 10 sec and then exhale in order to collect breath samples containing more than a 2% concentration of CO2. Samples were then analyzed by both methods. Results. The correlation analysis using values at 20 min after the administration of the study drug (433 points) was excellent with the regression equation of Y = 1.034x - 0.203; r = .996. The results of the UBiT-100 were available in 6 min, making the entire testing procedure less than 30 min. Conclusions. The UBiT-100 infrared analyzer provides a simple and accurate method of performing the urea breath test while the patient is still in the doctor's office.
引用
收藏
页码:49 / 53
页数:5
相关论文
共 13 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Nondispersive isotope-selective infrared spectroscopy: A new analytical method for C-13-urea breath tests [J].
Braden, B ;
Schafer, F ;
Caspary, WF ;
Lembcke, B .
SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1996, 31 (05) :442-445
[3]   ACCURACY OF INVASIVE AND NONINVASIVE TESTS TO DIAGNOSE HELICOBACTER-PYLORI INFECTION [J].
CUTLER, AF ;
HAVSTAD, S ;
MA, CK ;
BLASER, MJ ;
PEREZPEREZ, GI ;
SCHUBERT, TT .
GASTROENTEROLOGY, 1995, 109 (01) :136-141
[4]  
GRAHAM DY, 1987, LANCET, V1, P1174
[5]  
Kajiwara M., 1997, AM CLIN LAB, V16, P28
[6]  
Klein PD, 1996, AM J GASTROENTEROL, V91, P690
[7]   ISOTOPE-SELECTIVE NONDISPERSIVE INFRARED SPECTROMETRY FOR DETECTION OF HELICOBACTER-PYLORI INFECTION WITH C-13-UREA BREATH TEST [J].
KOLETZKO, S ;
HAISCH, M ;
SEEBOTH, I ;
BRADEN, B ;
HENGELS, K ;
KOLETZKO, B ;
HERING, P .
LANCET, 1995, 345 (8955) :961-962
[8]   SIMPLIFIED SINGLE SAMPLE C-13 UREA BREATH TEST FOR HELICOBACTER-PYLORI - COMPARISON WITH HISTOLOGY, CULTURE, AND ELISA SEROLOGY [J].
LOGAN, RPH ;
POLSON, RJ ;
MISIEWICZ, JJ ;
RAO, G ;
KARIM, NQ ;
NEWELL, D ;
JOHNSON, P ;
WADSWORTH, J ;
WALKER, MM ;
BARON, JH .
GUT, 1991, 32 (12) :1461-1464
[9]   LASER-BASED ANALYSIS OF CARBON-ISOTOPE RATIOS [J].
MURNICK, DE ;
PEER, BJ .
SCIENCE, 1994, 263 (5149) :945-947
[10]  
OHARA S, IN PRESS J GASTROENT