Quantitative measurement of male steroid hormones using automated on-line solid phase extraction-liquid chromatography-tandem mass spectrometry and comparison with radioimmunoassay

被引:41
作者
Chang, YC [1 ]
Li, CM
Li, LA
Jong, SB
Liao, PC
Chang, LW
机构
[1] Natl Hlth Res Inst, Div Environm Hlth & Occupat Med, Taipei, Taiwan
[2] Kaohsiung Med Univ, Dept Med Technol, Kaohsiung, Taiwan
[3] Natl Cheng Kung Univ, Grad Inst Environm & Occupat Hlth, Coll Med, Tainan 70101, Taiwan
关键词
D O I
10.1039/b210111b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A specific and sensitive method using high-performance liquid chromatography-tandem mass spectrometry (LC-MS-MS) equipped with automatic on-line solid-phase extraction device for the quantitative measurement of anabolic hormone residues, 4-androstene-3,17-dione, testosterone and dihydrotestosterone in cell culture medium was developed. Steroid content in cell culture medium was determined directly without an additional sample preparation step. Separation of analytes from polar endogenous compounds was carried out on an automatic column-switching device coupled with a C4-alkyl-diol silica restricted-access solid-phase extraction column. The lipophilic fraction containing anabolic hormone residues were back-flushed on to a conventional C-18 reversed-phase column for the final chromatography. The analyte was ionized in an ElectroSpray interface under positive ion mode before entering a quadrupole mass analyzer. The lowest points of calibration curves were 0.05 ng ml(-1) for 4-androstene-3,17-dione and testosterone, and 1 ng ml(-1) dihydrotestosterone, respectively. A comparison with results from radioimmunoassay (RIA) is also presented.
引用
收藏
页码:363 / 368
页数:6
相关论文
共 18 条
[1]  
Bean KA, 1997, J CHROMATOGR B, V690, P65, DOI 10.1016/S0378-4347(96)00403-3
[2]   Direct measurement of steroid sulfate and glucuronide conjugates with high-performance liquid chromatography mass spectrometry [J].
Bowers, LD ;
Sanaullah .
JOURNAL OF CHROMATOGRAPHY B, 1996, 687 (01) :61-68
[3]  
Croley TR, 2000, RAPID COMMUN MASS SP, V14, P1087
[4]   Measurement of steroid sex hormones in serum: a comparison of radioimmunoassay and mass spectrometry [J].
Dorgan, JF ;
Fears, TR ;
McMahon, RP ;
Friedman, LA ;
Patterson, BH ;
Greenhut, SF .
STEROIDS, 2002, 67 (3-4) :151-158
[5]   Quantitation of anabolic hormones and their metabolites in bovine serum and urine by liquid chromatography-tandem mass spectrometry [J].
Draisci, R ;
Palleschi, L ;
Ferretti, E ;
Lucentini, L ;
Cammarata, P .
JOURNAL OF CHROMATOGRAPHY A, 2000, 870 (1-2) :511-522
[6]   Column switching in high-performance liquid chromatography with tandem mass spectrometric detection for high-throughput preclinical pharmacokinetic studies [J].
Gao, VCX ;
Luo, WC ;
Ye, QP ;
Thoolen, M .
JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) :141-148
[7]  
HOUGHTON E, 1991, J REPROD FERTIL, P609
[8]   The quantification of endogenous steroids in bovine aqueous humour and vitreous humour using isotope dilution GC-NCI-MS [J].
Iqbal, Z ;
Midgely, JM ;
Watson, DG .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2001, 24 (04) :535-543
[9]   An evaluation of endocrine modulators: Implications for human health [J].
Juberg, DR .
ECOTOXICOLOGY AND ENVIRONMENTAL SAFETY, 2000, 45 (02) :93-105
[10]   An online automatic sample cleanup system for the quantitative detection of the benzene exposure biomarker S-phenylmercapturic acid in human urine by electrospray ionization tandem mass spectrometry [J].
Liao, PC ;
Li, CM ;
Lin, LC ;
Hung, CW ;
Shih, TS .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2002, 26 (04) :205-210