Use of the transient plane source technique for rapid multiple thermal property measurements

被引:26
作者
Boumaza, T [1 ]
Redgrove, J [1 ]
机构
[1] Natl Phys Lab, Thermophys Propert Grp, Teddington TW11 0LW, Middx, England
关键词
agar gel; alumina; ice; Perspex; silicone oil; thermal conductivity; transient technique; water;
D O I
10.1023/A:1022928206859
中图分类号
O414.1 [热力学];
学科分类号
摘要
This paper describes work at NPL to evaluate the capability of the transient plane source (TPS) technique using various sensor sizes and different types of materials that include solids (Perspex, alumina, extruded polystyrene, agar gel, and ice) and liquids (water and silicone oil). The aim of the present work is to investigate use of the TPS technique on materials where probe size, contact, and internal specimen convection are potentially important issues. Following validation of the technique on the NPL solid reference materials, measurements were carried out on ice using TPS and the NPL guarded hot plate (GHP) to illustrate the probe-to-sample thermal contact resistance issue. Measurements on silicone oil were compared to GHP and the NPL transient hot wire (THW) technique where the probe size/short times are crucial. In addition, measurements on water and agar gel were made to illustrate the influence of natural convection. Although the TPS is a multi-property technique, the focus of this work was on thermal conductivity.
引用
收藏
页码:501 / 512
页数:12
相关论文
共 12 条
[1]  
American Society of Heating Refrigerating and Air-Conditioning Engineers, 1997, ASHRAE HDB 1997 FUND
[2]  
CABANNES F, 1991, HIGH TEMPS HIGH PRES, V23, P589
[3]  
CORSAN JM, 1984, J PHYS E SCI INSTRUM, V17, P800, DOI 10.1088/0022-3735/17/9/018
[4]   THERMOPHYSICAL PROPERTIES OF ICE, SNOW, AND SEA ICE [J].
FUKUSAKO, S .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 1990, 11 (02) :353-372
[5]   THERMAL-PROPERTIES OF THIN INSULATING LAYERS USING PULSE TRANSIENT HOT STRIP MEASUREMENTS [J].
GUSTAFSSON, SE ;
CHOHAN, MA ;
AHMED, K ;
MAQSOOD, A .
JOURNAL OF APPLIED PHYSICS, 1984, 55 (09) :3348-3353
[6]   DETERMINATION OF THE THERMAL-CONDUCTIVITY TENSOR AND THE HEAT-CAPACITY OF INSULATING SOLIDS WITH THE TRANSIENT HOT-STRIP METHOD [J].
GUSTAFSSON, SE ;
KARAWACKI, E ;
KHAN, MN .
JOURNAL OF APPLIED PHYSICS, 1981, 52 (04) :2596-2600
[7]   TRANSIENT PLANE SOURCE TECHNIQUES FOR THERMAL-CONDUCTIVITY AND THERMAL-DIFFUSIVITY MEASUREMENTS OF SOLID MATERIALS [J].
GUSTAFSSON, SE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (03) :797-804
[8]  
GUSTAFSSON SE, 1994, P 22 INT THERM COND, P252
[9]  
KAYE GWC, 1995, TABLE PHYSICAL CHEM, P93
[10]   AN INSTRUMENT FOR THE MEASUREMENT OF THE THERMAL-CONDUCTIVITY OF LIQUIDS AT HIGH-TEMPERATURES [J].
POWELL, JS .