Initial experiments in high-performance liquid chromatographic method development .1. Use of a starting gradient run

被引:82
作者
Snyder, LR
Dolan, JW
机构
[1] LC Resources Inc., Orinda, CA 94563
基金
美国国家卫生研究院;
关键词
method development; gradient elution; mobile-phase composition; anxiolytics; pesticides; benzenes; phenazines; triazines; polynuclear aromatic hydrocarbons;
D O I
10.1016/0021-9673(95)00770-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A single gradient elution run with acetonitrile (B)-water (A) as mobile phase can be used to estimate preferred conditions for subsequent method development experiments based on reversed-phase high-performance liquid chromatography. For a broad range of sample types, that includes both very hydrophilic and hydrophobic compounds, it was found that isocratic retention is given by log k approximate to log k(w) - 4.2 phi, where phi = 0.01 %B. An initial gradient run allows values of log k(w) to be estimated for each compound in the sample, which then permits compound retention to be approximated as a function of either isocratic or gradient experimental conditions. The use of an initial gradient run in this way provides a rational basis for the subsequent development of a final HPLC method. Predictions based on this initial run can be used to (a) select between isocratic or gradient elution for further experiments and the final method, (b) choose a value of %B to achieve a desired value of k for the initial or final band in isocratic separation and/or (c) choose values of %B for the initial and final mobile phase in gradient elution. The present approach is based on a wide range of sample types and different reversed-phase columns; for this reason it is expected to be reasonably general and accurate. Errors in subsequent predictions based on an initial gradient run can also be estimated as a function of experimental conditions, allowing the selection of conditions for minimum error.
引用
收藏
页码:3 / 14
页数:12
相关论文
共 28 条