Specific determination of bromate and iodate in ozonized water by ion chromatography with postcolumn derivatization and inductively-coupled plasma mass spectrometry

被引:61
作者
Yamanaka, M [1 ]
Sakai, T [1 ]
Kumagai, H [1 ]
Inoue, Y [1 ]
机构
[1] Yokogawa Analyt Syst Inc, Applicat Dev Div, Musashino, Tokyo 180, Japan
关键词
water analysis; bromate; iodate; inorganic anions; halogen anions;
D O I
10.1016/S0021-9673(97)00518-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
eA specific determination for bromate, iodate and other halogen anions in drinking water by direct injection using ion chromatography (IC), with either inductively-coupled plasma mass spectrometry (ICP-MS) or postcolumn derivatization, is described. The advantages of ICP-MS as an element-selective detector was evaluated for bromate and iodate by considering the comparison with the postcolumn derivatization. Samples were directly injected into the IC column, and halogen anions were separated. The eluates were directly introduced into ICP-MS and detected at 79 and 127 u. The detection limit (S/N = 3) for bromate and iodate with injection of 0.5 ml were 0.45 mu g/l and 0.034 mu g/l, respectively. IC combined with ICP-MS was applied to the simultaneous determination of bromate, bromide and other halogen anions in raw and ozonized water. Good agreement was obtained for the determined values by IC-ICP MS and postcolumn derivatization. Furthermore, several bromine species different from bromate or bromide were detected by IC-ICP-MS. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 17 条
[1]   Determination of bromate in drinking waters by ion chromatography with inductively coupled plasma mass spectrometric detection [J].
Creed, JT ;
Magnuson, ML ;
Pfaff, JD ;
Brockhoff, C .
JOURNAL OF CHROMATOGRAPHY A, 1996, 753 (02) :261-267
[2]   PRACTICAL APPLICATIONS OF ELEMENT-SPECIFIC DETECTION BY INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROSCOPY AND INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY TO TON CHROMATOGRAPHY OF FOODS [J].
HEITKEMPER, DT ;
KAINE, LA ;
JACKSON, DS ;
WOLNIK, KA .
JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) :101-108
[3]   SIMULTANEOUS DETERMINATION OF CL-, BR- AND I- IN WATER SAMPLES BY ISOTOPE-DILUTION MASS-SPECTROMETRY [J].
HEUMANN, KG ;
SEEWALD, H .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 320 (05) :493-497
[4]   ELEMENTAL SPECIATION WITH LIQUID-CHROMATOGRAPHY INDUCTIVELY-COUPLED PLASMA ISOTOPE-DILUTION MASS-SPECTROMETRY [J].
HEUMANN, KG ;
ROTTMANN, L ;
VOGL, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) :1351-1355
[5]  
INOUE Y, IN PRESS ANAL CHIM A
[6]   ION CHROMATOGRAPHY OF INORGANIC IODINE SPECIES USING C-18 REVERSED-PHASE COLUMNS COATED WITH CETYLTRIMETHYLAMMONIUM [J].
ITO, K ;
SHOTO, E ;
SUNAHARA, H .
JOURNAL OF CHROMATOGRAPHY, 1991, 549 (1-2) :265-272
[7]   TRACE-LEVEL DETERMINATION OF BROMATE IN OZONATED DRINKING-WATER USING ION CHROMATOGRAPHY [J].
JOYCE, RJ ;
DHILLON, HS .
JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) :165-171
[8]  
KRUITHOF JC, 1995, WATER SUPP, V13, P93
[9]  
KRUITHOF JC, 1995, WATER SUPP, V13, P117
[10]  
KRUITHOF JC, 1992, WATER SUPPLY, V11, P121