The platelet count accuracy of platelet concentrates obtained by using automated analysis is influenced by instrument bias and activated platelet components

被引:22
作者
Hervig, T [1 ]
Haugen, T
Liseth, K
Kjeldsen-Kragh, J
Scott, CS
Johannessen, B
机构
[1] Haukeland Hosp, Blood Bank, N-5021 Bergen, Norway
[2] Ullevaal Univ Hosp, Dept Immunol & Transfus Med, Oslo, Norway
[3] Abbott Diagnost GmbH, Wiesbaden, Germany
[4] Haukeland Hosp, Dept Clin Biochem, N-5021 Bergen, Norway
关键词
activated platelets; immunoplatelet analysis; platelet concentrates; platelet counting;
D O I
10.1111/j.1423-0410.2004.00557.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives The blood platelet content (in numbers) of platelet concentrates is required for production quality control and to predict clinical responses. Materials and Methods This study compared the performance of automated counting from impedance and optical instruments to data from immunoplatelet reference analysis. Results All methods showed good linearity with evidence of significant instrument-specific deviations from the line of agreement. Relational formulae largely corrected bias, but did not resolve platelet count variability. A second confounding factor, related to the proportion of small (activated) platelets, was also shown to contribute to intermethod discrepancies. Conclusions Blood processing centres should establish correction factors for each instrument compared to reference methods, such as the immunoplatelet count.
引用
收藏
页码:196 / 203
页数:8
相关论文
共 14 条
[1]   Assessment of platelet activation in several different anticoagulants by the Advia 120 Hematology System, fluorescence flow cytometry, and electron microscopy [J].
Ahnadi, CE ;
Chapman, ES ;
Lépine, M ;
Okrongly, D ;
Pujol-Moix, N ;
Hernández, A ;
Boughrassa, F ;
Grant, AM .
THROMBOSIS AND HAEMOSTASIS, 2003, 90 (05) :940-948
[2]   Evaluation of a CD61 MoAb method for enumeration of platelets in thrombocytopenic patients and its impact on the transfusion decision-making process [J].
Arroyo, JL ;
García-Marcos, MA ;
López, A ;
Menéndez, P ;
Tabernero, MD ;
Sánchez-Abarca, LI ;
Avila-Zarza, C ;
San Miguel, JF ;
Orfao, A .
TRANSFUSION, 2001, 41 (10) :1212-1216
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
BODE AP, 1991, BLOOD, V77, P887
[5]   Corrected count increment and percent platelet recovery as measures of posttransfusion platelet response: problems and a solution [J].
Davis, KB ;
Slichter, SJ ;
Corash, L .
TRANSFUSION, 1999, 39 (06) :586-592
[6]   A rapid and accurate closed-tube immunoassay for platelets on an automated hematology analyzer [J].
Gill, JE ;
Davis, KA ;
Cowart, WJ ;
Nepacena, FU ;
Kim, YR .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 114 (01) :47-56
[7]  
Harrison P, 2000, BRIT J HAEMATOL, V108, P228
[8]   Standardisation of platelet counting accuracy in blood banks by reference to an automated immunoplatelet procedure: comparative evaluation of Cell-Dyn CD4000 impedance and optical platelet counts [J].
Johannessen, B ;
Haugen, T ;
Scott, CS .
TRANSFUSION AND APHERESIS SCIENCE, 2001, 25 (02) :93-106
[9]  
Klee G, 2001, AM J CLIN PATHOL, V115, P460
[10]   Automated CD61 immunoplatelet analysis of thrombocytopenic samples [J].
Kunz, D ;
Kunz, WS ;
Scott, CS ;
Gressner, AM .
BRITISH JOURNAL OF HAEMATOLOGY, 2001, 112 (03) :584-592