Lithium-ion batteries for hearing aid applications II. Pulse discharge and safety tests

被引:15
作者
Passerini, S [1 ]
Coustier, F [1 ]
Owens, BB [1 ]
机构
[1] Res Int, Woodinville, WA 98072 USA
关键词
Li-ion button cell; hearing aid cell; hearing aid; safety;
D O I
10.1016/S0378-7753(00)00396-7
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Rechargeable lithium-ion batteries were designed to meet the power requirements of hearing aid devices (HADs). The batteries were designed in a 312-button cell size, compatible with existing hearing aids. The batteries were tested to evaluate the design and the electrochemical performance, as they relate to a typical hearing aid application. The present report covers the pulse capabilities, cycle life and preliminary safety tests. The results are compared with other battery chemistries: secondary lithium-alloy and nickel-metal hydride batteries and primary Zn-air batteries. The cell AC impedance was stable over the frequency range between 1 and 50 kHz, ranging between 5 Omega at the higher frequency and 12 Omega at the lower extreme. Pulse tests were consistent with these values, as the cells were capable of providing a series of 100 mA pulses of 10-s duration. The safety tests suggest that the design is intrinsically safe with respect to the most common types of abuse conditions. (C) 2000 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:144 / 152
页数:9
相关论文