Investigation of a solid-state detector for advanced computed tomography

被引:32
作者
Hsieh, J [1 ]
Gurmen, OE
King, KF
机构
[1] GE, Med Syst, Appl Sci Lab, Milwaukee, WI 53201 USA
[2] GE, Med Syst, Detector Engn, Milwaukee, WI 53201 USA
关键词
computed tomography; primary speed and afterglow; recursive correction; solid-state detector;
D O I
10.1109/42.887840
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Utilization of solid-state detectors for computed tomography (CT) has been the focus of many studies. Previous phantom and clinical experiments have shown that one of the important performance parameters for the solid-state detector is the primary speed and afterglow, In this paper, we present a detailed investigation on the signal decay characteristics of the HiLight (GE Medical Systems, Milwaukee, WI) scintillating detector, The detector primary speed and afterglow are modeled by a multiexponential function and fully characterized by a set of time constants and relative strengths. The sensitivity of these parameters to X-ray photon energy, detector aging, and radiation exposure is then established and analyzed. No statistically significant variation is observed in these parameters due to changes in the above external variables. The impact of various decay time constants on CT image quality, such as spatial resolution, noise, and artifacts, is subsequently illustrated with computer simulations and phantom experiments, Finally, an algorithmic correction scheme is derived to compensate for detector afterglow The correction scheme employs a recursive filter to remove adverse effects of the detector decay on image quality. Experimental results have shown that the correction scheme successfully restores system spatial resolution, produces a more homogeneous noise pattern, and eliminates ring-band image artifacts due to detector afterglow. The effectiveness and robustness of the correction scheme are demonstrated by extensive phantom and clinical experiments.
引用
收藏
页码:930 / 940
页数:11
相关论文
共 14 条
  • [1] CUSANO D, 1988, Patent No. 4747973
  • [2] CUSANO D, 1983, Patent No. 4421671
  • [3] FOX SH, 1997, GEN ELECT CO PUBL
  • [4] NEW SCINTILLATORS FOR X-RAY COMPUTED-TOMOGRAPHY
    GRABMAIER, BC
    ROSSNER, W
    [J]. NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1993, 21 (01): : 43 - 45
  • [5] Greskovich C., 1984, US patent, Patent No. 4466930
  • [6] Greskovich C D, 1985, US Patent, Patent No. [4 525 628, 4525628]
  • [7] GRESKOVICH CD, 1992, AM CERAM SOC BULL, V71, P1120
  • [8] Hsieh J., 1993, U.S. patent, Patent No. [5249123 A, 5249123]
  • [9] JOHNS HE, 1983, PHYSICS RADIOLOGY
  • [10] King K. F., 1993, U.S. Patent, Patent No. [5 265 013, 5265013]