Prevalence and clinical impact of stent fractures after femoropopliteal stenting

被引:781
作者
Scheinert, D [1 ]
Scheinert, S [1 ]
Sax, J [1 ]
Piorkowski, C [1 ]
Bräunlich, S [1 ]
Ulrich, M [1 ]
Biamino, G [1 ]
Schmidt, A [1 ]
机构
[1] Univ Leipzig, Ctr Heart, Dept Med Cardiol, D-04289 Leipzig, Germany
关键词
D O I
10.1016/j.jacc.2004.11.026
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The aim of this study was to investigate the occurrence and the clinical impact of stent fractures after femoropopliteal stenting. BACKGROUND The development of femoral stent fractures has recently been described; however, there are no data about the frequency and the clinical relevance. METHODS A systematic X-ray screening for stent fractures was performed in 93 patients. In total, 121 legs treated by implantation of self-expanding nitinol stents were investigated after a mean follow-up time of 10.7 months. The mean length of the stented segment was 15.7 cm. RESULTS Overall, stent fractures were detected in 45 of 121 treated legs (37.2%). In a stent-based analysis, 64 of 261 stents (24.5%) showed fractures, which were classified as minor (single strut fracture) in 31 cases (48.4%), moderate (fracture of >1 strut) in 17 cases (26.6%), and severe (complete separation of stent segments) in 16 cases (25.0%). Fracture rates were 13.2% for stented length less than or equal to8 cm, 42.4% for stented length >8 to 16 cm and 52.0% for stented length >16 cm. In 21 cases (32.8%) there was a restenosis of >50% diameter reduction at the site of stent fracture. In 22 cases (34.4%) with stent fracture there was a total stent reocclusion. According to Kaplan-Meier estimates, the primary patency rate at 12 months was significantly lower for patients with stent fractures (41.1% vs. 84.3%, p < 0.0001). CONCLUSIONS There is a considerable risk of stent fractures after long segment femoral artery stenting, which is associated with a higher in-stent restenosis and reocclusion rate. (C) 2005 by the American College of Cardiology Foundation.
引用
收藏
页码:312 / 315
页数:4
相关论文
共 12 条
  • [1] Fracture of popliteal artery stents
    Babalik, E
    Gülbaran, M
    Gürmen, T
    Öztürk, S
    [J]. CIRCULATION JOURNAL, 2003, 67 (07) : 643 - 645
  • [2] Angioplasty and stent placement in chronic occlusion of the superficial femoral artery: Technique and results
    Conroy, RM
    Gordon, IL
    Tobis, JM
    Hiro, T
    Kasaoka, S
    Stemmer, EA
    Wilson, SE
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (08) : 1009 - 1020
  • [3] Sirolimus-eluting stents for the treatment of obstructive superficial femoral artery disease
    Duda, SH
    Pusich, B
    Richter, G
    Landwehr, P
    Oliva, VL
    Tielbeek, A
    Wiesinger, B
    Hak, JB
    Tielemans, H
    Ziemer, G
    Cristea, E
    Lansky, A
    Bérégi, JP
    [J]. CIRCULATION, 2002, 106 (12) : 1505 - 1509
  • [4] High incidence of restenosis/reocclusion of stents in the percutaneous treatment of long-segment superficial femoral artery disease after suboptimal angioplasty
    Gray, BH
    Sullivan, TM
    Childs, MB
    Young, JR
    Olin, JW
    [J]. JOURNAL OF VASCULAR SURGERY, 1997, 25 (01) : 74 - 83
  • [5] Mechanical failure of prosthetic human implants: A 10-year experience with aortic stent graft devices
    Jacobs, TS
    Won, J
    Gravereaux, EC
    Faries, PL
    Morrissey, N
    Teodorescu, VJ
    Hollier, LH
    Marin, ML
    [J]. JOURNAL OF VASCULAR SURGERY, 2003, 37 (01) : 16 - 26
  • [6] Endovascular placement of self-expanding nitinol coil stents for the treatment of femoropopliteal obstructive disease
    Jahnke, T
    Voshage, G
    Müller-Hülsbeck, S
    Grimm, J
    Heller, M
    Brossmann, J
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (03) : 257 - 266
  • [7] Longitudinal Stent fracture 11 months after implantation in the left pulmonary artery and successful management by a stent-in-stent maneuver
    Knirsch, W
    Haas, NA
    Lewin, MAG
    Uhlemann, F
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 58 (01) : 116 - 118
  • [8] Treatment of complex arteriosclerotic lesions with nitinol stents in the superficial femoral and popliteal arteries: A midterm follow-up
    Lugmayr, HF
    Holzer, H
    Kastner, M
    Riedelsberger, H
    Auterith, A
    [J]. RADIOLOGY, 2002, 222 (01) : 37 - 43
  • [9] INTRAVASCULAR STENTS FOR MANAGEMENT OF PULMONARY-ARTERY AND RIGHT-VENTRICULAR OUTFLOW OBSTRUCTION
    NAKANISHI, T
    KONDOH, C
    NISHIKAWA, T
    SATOMI, G
    NAKAZAWA, M
    IMAI, Y
    MOMMA, K
    [J]. HEART AND VESSELS, 1994, 9 (01) : 40 - 48
  • [10] Perry Stanton B., 1993, Journal of the American College of Cardiology, V21, p261A