Improvement in high-resolution ECG analysis by interpolation before time alignment

被引:3
作者
Laguna, P [1 ]
Simon, B [1 ]
Sornmo, L [1 ]
机构
[1] Univ Zaragoza, Ctr Politecn Super, Dept Ingn Elect & Comunicaciones, Zaragoza 50015, Spain
来源
COMPUTERS IN CARDIOLOGY 1997, VOL 24 | 1997年 / 24卷
关键词
D O I
10.1109/CIC.1997.648125
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In this work we study the use of interpolation prior to time alignment and averaging in the analysis of high resolution ECGs. The resolution of time alignment is limited by the sampling interval and interpolation makes the alignment error reduced. Five alignment methods were used in combination with white noise and 50/60 Hz interference. The results were quantified in terms of power of the deviation signal, the deviation of the misalignment error and the energy of the error signal. A considerable decrease of the deviation signal power was noted when the sampling rate was increased from 1 kHz to 4 kHz. Modest improvements were only obtained when the sampling rate was further increased, making 4 kHz suitable sampling rate for high frequency studies. In actual and simulated ECG records it was found that the filtered QRS duration index increased only 4 ms whereas RMS40 and LAS40 remained unchanged.
引用
收藏
页码:617 / 620
页数:4