Fully adaptive finite element based tomography using tetrahedral dual-meshing for fluorescence enhanced optical imaging in tissue

被引:49
作者
Lee, Jae Hoon [1 ]
Joshi, Amit [1 ]
Sevick-Muraca, Eva M. [1 ]
机构
[1] Baylor Coll Med, Dept Radiol, Houston, TX 77030 USA
关键词
D O I
10.1364/OE.15.006955
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have developed fluorescence enhanced optical tomography based upon fully adaptive finite element method (FEM) using tetrahedral dual-meshing wherein one of the two meshes discretizes the forward variables and the other discretizes the unknown parameters to be estimated. We used the 8-subtetrahedron subdivision scheme to create the nested dual-mesh in which each are independently refined. However, two tetrahedrons from the two different meshes pose an intersection problem that needs to be resolved in order to find the common regions that the forward variables (the fluorescent diffuse photon fluence fields) and the parameter estimates (the fluorescent absorption coefficients) can be mutually assigned. Using an efficient intersection algorithm in the nested tetrahedral environments previously developed by the authors, we demonstrate fully adaptive tomography using a posteriori error estimates. Performing the iterative reconstructions using the simulated boundary measurement data, we demonstrate that small fluorescent targets embedded in the breast simulating phantom in point illumination/detection geometry can be resolved at reasonable computational cost. (C) 2007 Optical Society of America.
引用
收藏
页码:6955 / 6975
页数:21
相关论文
共 14 条
[1]   Three-dimensional optical tomography: resolution in small-object imaging [J].
Dehghani, H ;
Pogue, BW ;
Jiang, SD ;
Brooksby, B ;
Paulsen, KD .
APPLIED OPTICS, 2003, 42 (16) :3117-3128
[2]   A comparison of exact and approximate adjoint sensitivities in fluorescence tomography [J].
Eppstein, MJ ;
Fedele, F ;
Laible, J ;
Zhang, C ;
Godavarty, A ;
Sevick-Muraca, EM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (10) :1215-1223
[3]   Dual-mesh optical tomography reconstruction method with a depth correction that uses a priori ultrasound information [J].
Huang, MM ;
Zhu, Q .
APPLIED OPTICS, 2004, 43 (08) :1654-1662
[4]   Frequency-domain near-infrared photo diffusion imaging: Initial evaluation in multitarget tissuelike phantoms [J].
Jiang, HB ;
Paulsen, KD ;
Osterberg, UL ;
Patterson, MS .
MEDICAL PHYSICS, 1998, 25 (02) :183-193
[5]   Adaptive finite element based tomography for fluorescence optical imaging in tissue [J].
Joshi, A ;
Bangerth, W ;
Sevick-Muraca, EM .
OPTICS EXPRESS, 2004, 12 (22) :5402-5417
[6]   Non-contact fluorescence optical tomography with scanning patterned illumination [J].
Joshi, Amit ;
Bangerth, Wolfgang ;
Sevick-Muraca, Eva M. .
OPTICS EXPRESS, 2006, 14 (14) :6516-6534
[7]   A-POSTERIORI ERROR ANALYSIS AND ADAPTIVE PROCESSES IN THE FINITE-ELEMENT METHOD .1. ERROR ANALYSIS [J].
KELLY, DW ;
GAGO, JPDR ;
ZIENKIEWICZ, OC ;
BABUSKA, I .
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN ENGINEERING, 1983, 19 (11) :1593-1619
[8]  
LEE JH, UNPUB FAST INTERSECT
[9]   Quality local refinement of tetrahedral meshes based on 8-subtetrahedron subdivision [J].
Liu, AW ;
Joe, B .
MATHEMATICS OF COMPUTATION, 1996, 65 (215) :1183-1200
[10]   A DUAL MESH SCHEME FOR FINITE-ELEMENT BASED RECONSTRUCTION ALGORITHMS [J].
PAULSEN, KD ;
MEANEY, PM ;
MOSKOWITZ, MJ ;
SULLIVAN, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (03) :504-514