Wear of ultra-high molecular weight polyethylene and polytetrafluoroethylene in a hip simulator - A dose-response study of protein concentration

被引:34
作者
Good, VD
Clarke, IC
Gustafson, GA
Downs, B
Anissian, L
Sorensen, K
机构
[1] Loma Linda Univ, Med Ctr, Howard & Irene Peterson Tribol Lab, Dept Orthopaed Surg, San Bernardino, CA 92408 USA
[2] Univ Strathclyde, Glasgow, Lanark, Scotland
[3] Karolinska Univ Hosp, Stockholm, Sweden
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 2000年 / 71卷 / 04期
关键词
D O I
10.1080/000164700317393358
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Charnley's laboratory wear studies of non-gamma sterilized polyletrafluoroethylene (PTFE) and polyethylene (PE) found that the PTFE to PE wear-rate ratio of 250:1 was much higher than the in-vivo wear ratio of 20:1. Tests of PTFE and PE in our laboratory showed a wear ratio of 150:1, using bovine serum as the lubricant and 190:1 with water as the lubricant. Our hypothesis was that the wear-rates of PTFE and PE cup materials were related to the concentration of protein in the serum. We studied the wear behavior of PTFE and PE cups in varied protein concentrations, using 4 femoral head sizes to validate the clinical range reported by Charnley, The PTFE wear-rates increased with increasing protein concentration and conversely, PE wear-rates decreased with increasing protein concentration. This inverse relationship made it possible to bring the wear ratio closer to the desired clinical wear ratio. We found that the clinically relevant PTFE/PE wear ratio corresponded to 3-10 mg/mL of protein in bovine serum.
引用
收藏
页码:365 / 369
页数:5
相关论文
共 36 条
[1]  
ALTMAN RD, 1984, OSTEOARTHRITIS DIAGN, P185
[2]  
*ASTM COMM, F042210 ASTM COMM ME
[3]  
BENNETT AP, 1996, UHMWPE 42 ANN M ORTH, P472
[4]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[5]  
BRAGDON CR, 1998, COMPARISON VOLUMETRI, P53
[6]  
CALLAGHAN JJ, 1995, CLIN ORTHOP RELAT R, P14
[7]   OPTIMUM SIZE OF PROSTHETIC HEADS IN RELATION TO WEAR OF PLASTIC SOCKETS IN TOTAL REPLACEMENT OF HIP [J].
CHARNLEY, J ;
KAMANGAR, A ;
LONGFIELD, MD .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :31-+
[8]  
CHARNLEY J, 1976, PLAST RUBBER INT, V1, P59
[9]   Charnley wear model for validation of hip simulators - Ball diameter versus polytetrafluoroethylene and polyethylene wear [J].
Clarke, IC ;
Good, V ;
Anissian, L ;
Gustafson, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :25-36
[10]  
CLARKE IC, 1995, T 2 COMB ORTH RES SO, P204