Speciation of chromium by direct coupling of ion exchange chromatography with inductively coupled plasma mass spectrometry

被引:102
作者
Barnowski, C
Jakubowski, N
Stuewer, D
Broekaert, JAC
机构
[1] INST SPEKTROCHEM & ANGEW SPEKTROSKOPIE,D-44013 DORTMUND,GERMANY
[2] UNIV DORTMUND,D-44221 DORTMUND,GERMANY
关键词
chromium; speciation; ion exchange chromatography; inductively coupled plasma; mass spectrometry; hydraulic high pressure nebulization;
D O I
10.1039/a702120h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
inductively coupled plasma mass spectrometry (ICP-MS) for the element speciation of chromium. Hydraulic high pressure nebulization was used for sample introduction with high aerosol efficiency. It was the particular aim of this work to exclusively use nitric acid for elution in order to reduce interferences in ICP-MS. Therefore, an anion exchange precolumn IonPac-AG5 (Dionex, Idstein, Germany) with a length of 50 mm and an id of 4 mm has been operated as a mixed-mode column, which Shows cation exchange capabilities too. The column was operated at a flow rate of 1.2 ml min(-1) and samples were injected by use of a sample loop with a volume of 100 mu l. Both Cr-VI anions and Cr-III cations could be retained on the column, and optimum results concerning. chromatographic and mass spectrometric conditions were finally achieved with discontinuous elution in two steps by injection of 0.3 M (pH 0.5) HNO3 for Cr-VI add subsequently of 1.0 M (pH 0) HNO3 for Crm, Detection limits just above 0.1 mu g l(-1) could be realized.
引用
收藏
页码:1155 / 1161
页数:7
相关论文
共 27 条