Noninvasive measurement of oxygenation of hemoglobin by direct transmission of near-infrared energy (700-1000 nm) from an electronically tuned Ti:sapphire laser driven by a dual radio-frequency driving method

被引:21
作者
Sato, H
Wada, S
Ling, MY
Tashiro, H
机构
[1] Inst Phys & Chem Res, Dept Res Fundamentals Technol, Div Fundamental Technol Dev, Wako, Saitama 3510198, Japan
[2] Photon Tuning, Aoba Ku, Sendai, Miyagi 9893201, Japan
关键词
tunable laser; NIR; medical; hemoglobin; non-invasive analysis;
D O I
10.1366/0003702001950977
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have applied an electronically tuned Ti:sapphire (ETT) laser controlled with a dual radio-frequency driving (DRD) method to noninvasive measurements of the transmission spectra of a whole human hand in which the bands due to hemoglobin and lipid were identified. Spectra of a human hand that is more than 3 cm thick were obtained in the spectral range from 700 to 1000 nm, allowing one to probe the oxygenation of hemoglobin (Hb). Our study demonstrates that an ETT laser may be used as a potential light source For near-infrared (NIR) spectroscopy in biomedical applications. For the acquisition of NIR transmission spectra of Hb with a good signal-to-noise ratio, the laser emission was programmed electronically to fill up strong absorption bands due to water, and baseline undulation due to detection sensitivity and light scattering, The present study required more stable control of the ETT laser, so we developed the DRD method, achieving high reliability in wavelength (<0.1 nm) and in intensity (< +/- 2% per average in 10 pulses).
引用
收藏
页码:1163 / 1167
页数:5
相关论文
共 16 条
[1]   High-speed optical parametric oscillator pumped with an electronically tuned Ti:Sapphire laser [J].
Akagawa, K ;
Wada, S ;
Tashiro, H .
APPLIED PHYSICS LETTERS, 1997, 70 (10) :1213-1215
[2]  
Cote Gerard L., 1997, Journal of Clinical Engineering, V22, P253
[3]   Multivariate calibration for near-infrared spectroscopic assays of blood substrates in human plasma based on variable selection using PLS-regression vector choices [J].
Heise, HM ;
Bittner, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1998, 362 (01) :141-147
[4]   Non-invasive monitoring of metabolites using near infrared spectroscopy: State of the art [J].
Heise, HM .
HORMONE AND METABOLIC RESEARCH, 1996, 28 (10) :527-534
[5]  
Kakefuda T, 1996, TRANSPLANT P, V28, P1792
[6]  
KANASHIRO M, 1998, THER RES, V19, P2119
[7]  
Khalil OS, 1999, CLIN CHEM, V45, P165
[8]   ABSOLUTE QUANTIFICATION OF DEOXYHEMOGLOBIN CONCENTRATION IN TISSUE NEAR-INFRARED SPECTROSCOPY [J].
MATCHER, SJ ;
COOPER, CE .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) :1295-1312
[9]   USE OF THE WATER-ABSORPTION SPECTRUM TO QUANTIFY TISSUE CHROMOPHORE CONCENTRATION CHANGES IN NEAR-INFRARED SPECTROSCOPY [J].
MATCHER, SJ ;
COPE, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (01) :177-196
[10]   SIMULTANEOUS EVALUATION OF NITRIC-OXIDE SYNTHESIS AND TISSUE OXYGENATION IN RAT-LIVER ALLOGRAFT-REJECTION USING NEAR-INFRARED SPECTROSCOPY [J].
OHDAN, H ;
FUKUDA, Y ;
SUZUKI, S ;
AMEMIYA, H ;
DOHI, K .
TRANSPLANTATION, 1995, 60 (06) :530-535