A novel contact angle measurement technique by analysis of capillary rise profile around a cylinder (ACRPAC)

被引:32
作者
Gu, YG [1 ]
Li, DQ [1 ]
Cheng, P [1 ]
机构
[1] UNIV ALBERTA,DEPT MECH ENGN,EDMONTON,AB T6G 2G8,CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
capillary rise; contact angle; fibre; line tension;
D O I
10.1016/S0927-7757(96)03853-8
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
A new experimental technique and its computational scheme to determine the contact angles of capillary rise profiles around a cylinder are presented. In the experiment, a carefully coated conic glass cylinder was inserted vertically and slowly into a tested liquid. Then the precise image of the partial capillary rise profile of the liquid around the conic cylinder was acquired and digitized by applying computer image processing and analysis techniques. From the digitized profiles of the liquid-vapour interface and the conic cylinder, the local inclination angle beta and the local radius R-c, of the conic cylinder at the three-phase contact circle were calculated directly. Furthermore, an objective function was constructed, which expresses the discrepancy between the physically observed capillary rise profile and the theoretically predicted curve, i.e. the curve representing a solution of the Laplace equation of capillarity. The contact angle of the capillary rise profile on the conic cylinder was used as an adjustable parameter in optimizing the objective function and determined once the minimum objective function had been achieved. The accuracy of the measured contact angles is approximately 0.1 degrees. In addition to the local gravity, the densities of the liquid and vapour phases and the liquid-vapour surface tension, the input requirement is the digital information of the partial capillary rise profile which is provided by a specially designed computer image analysis program. This method was tested by measuring contact angles of four n-alkane liquids around cylindrical glass fibres coated with FC725. The measured contact angles are in very good agreement with those determined by the Wilhelmy plate technique. Finally, the present technique was also applied to study the dependence of contact angles on the geometry of the conic cylinder, i.e. on cos beta/R-c. Contact angles of the four n-alkane liquids on a conic glass cylinder coated with FC725 were measured at different positions along the cylinder. The results were interpreted in terms of the line tension effect. The calculated line tensions were positive and of the order of 1 mu J m(-1), which is consistent with the published data for similar solid-liquid systems obtained by using the sessile drop method. In particular, the contact angle without Line tension effect theta(infinity) for a given solid-liquid system can be measured directly by this method. The validity of the derived contact angle theta(infinity) and line tension sigma was also confirmed by means of the axisymmetric drop shape analysis (ADSA) technique. This novel technique is particularly suitable to the study of the wetting and spreading phenomena of a liquid on fibres, as most fibre surfaces are rough and their shapes may deviate considerably from those of right circular cylinders. A general user-oriented computer program to implement the technique was developed.
引用
收藏
页码:135 / 149
页数:15
相关论文
共 36 条
[1]  
ADAMSON AW, 1982, PHYSICAL CHEM SURFAC
[2]  
Ambwani D.S., 1979, Surface and Colloid Science, P93, DOI DOI 10.1007/978-1-4615-7969-4_3
[3]   GENERALIZATION OF CLASSICAL-THEORY OF CAPILLARITY [J].
BORUVKA, L ;
NEUMANN, AW .
JOURNAL OF CHEMICAL PHYSICS, 1977, 66 (12) :5464-5476
[4]   AUTOMATION OF THE CAPILLARY RISE TECHNIQUE FOR MEASURING CONTACT ANGLES [J].
BUDZIAK, CJ ;
NEUMANN, AW .
COLLOIDS AND SURFACES, 1990, 43 (2-4) :279-293
[5]   ACCURATE MEASUREMENT OF CONTACT-ANGLE, PHASE CONTACT AREAS, DROP VOLUME, AND LAPLACE EXCESS PRESSURE IN DROP-ON-FIBER SYSTEMS [J].
CARROLL, BJ .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1976, 57 (03) :488-495
[6]   AUTOMATION OF AXISYMMETRIC DROP SHAPE-ANALYSIS FOR MEASUREMENT OF INTERFACIAL-TENSIONS AND CONTACT ANGLES [J].
CHENG, P ;
LI, D ;
BORUVKA, L ;
ROTENBERG, Y ;
NEUMANN, AW .
COLLOIDS AND SURFACES, 1990, 43 (2-4) :151-167
[7]   CORRELATION OF LINE TENSION AND SOLID-LIQUID INTERFACIAL-TENSION FROM THE MEASUREMENT OF DROP SIZE DEPENDENCE OF CONTACT ANGLES [J].
DUNCAN, D ;
LI, D ;
GAYDOS, J ;
NEUMANN, AW .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1995, 169 (02) :256-261
[8]   THE DEPENDENCE OF CONTACT ANGLES ON DROP SIZE AND LINE TENSION [J].
GAYDOS, J ;
NEUMANN, AW .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1987, 120 (01) :76-86
[9]  
Gibbs J.W., 1961, SCI PAPERS JW GIBBS, V1, P288
[10]   EFFECT OF DROP SIZE ON CONTACT-ANGLE [J].
GOOD, RJ ;
KOO, MN .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1979, 71 (02) :283-292