A simple technique for measuring the infrared emissivity of black-body radiators

被引:14
作者
Ballico, M [1 ]
机构
[1] CSIRO, Div Appl Phys, Natl Measurement Lab, Lindfield, NSW 2070, Australia
关键词
D O I
10.1088/0026-1394/37/4/5
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Black-body radiators, generally consisting of a nearly isothermal cavity with a surface of high intrinsic emissivity, are used in radiation thermometry for the transfer of a temperature scale to a radiation thermometer from either a contact thermometer or another radiation thermometer. The emissivity of these black bodies is typically greater than 0.99, and is usually estimated by calculation. In the commonly used 7 mum to 14 mum band, several factors conspire to make this calculation less satisfactory, and a technique is required to confirm the calculation by direct measurement of the cavity emissivity. A simple technique is presented for determining the average normal-hemispherical reflectance, and hence emissivity, of a cavity radiator over this wavelength region. An uncertainty of 0.0002 (k = 2) in emissivity is achieved, and experimentally measured and calculated emissivities are presented for the cavities used at the National Measurement Laboratory (NML, Australia) for radiation thermometer calibration. An assessment is also made of the reflectance of an "ice-cavity" radiator.
引用
收藏
页码:295 / 300
页数:6
相关论文
共 12 条
[1]   Modelling of the effective emissivity of a graphite tube black body [J].
Ballico, MJ .
METROLOGIA, 1996, 32 (04) :259-265
[2]  
BATTUELLO, 1999, P TEMPMEKO 99 7 IMEK, P601
[3]   EMISSIVITIES OF DIFFUSE CAVITIES .2. ISOTHERMAL AND NONISOTHERMAL CYLINDRO-CONES [J].
BEDFORD, RE ;
MA, CK .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1975, 65 (05) :565-572
[4]   DIFFRACTION LOSSES IN RADIOMETRY AND PHOTOMETRY [J].
BLEVIN, WR .
METROLOGIA, 1970, 6 (02) :39-&
[5]  
DEWITT DP, 1988, THEORY PRACTICE RAD, P653
[6]   Effective emissivities of black-body cavities with grooved cylinders [J].
Ishii, J ;
Kobayashi, M ;
Sakuma, F .
METROLOGIA, 1998, 35 (03) :175-180
[7]   REALIZATION OF IPTS68 ABOVE 1064.43 DEGREE C USING NSL PHOTOELECTRIC PYROMETER [J].
JONES, TP ;
TAPPING, J .
METROLOGIA, 1972, 8 (01) :4-+
[8]  
MACHIN G, 1999, P TEMPMEKO 99 7 INT, P687
[10]  
Palik ED, 1991, HDB OPTICAL CONSTANT