Assessment of heating effects in skin during continuous wave near infrared spectroscopy

被引:56
作者
Ito, Y
Kennan, RP
Watanabe, E
Koizumi, H [1 ]
机构
[1] Univ Tokyo, Grad Sch Arts & Sci, Meguro Ku, Tokyo, Japan
[2] Hitachi Cent Res Lab, Kokubunji, Tokyo 8158601, Japan
[3] Yale Univ, Sch Med, Dept Diagnost Radiol, New Haven, CT 06511 USA
[4] Tokyo Metropolitan Police Hosp, Dept Neurosurg, Chiyoda Ku, Tokyo 1028161, Japan
[5] Hitachi Adv Res Lab, Hatoyama, Saitama 3500395, Japan
关键词
optical imaging; near-infrared spectroscopy; laser heating;
D O I
10.1117/1.1287730
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Near infrared spectroscopy is an increasingly important tool for the investigation of human brain function, however, to date there have been few systematic evaluations of accompanying thermal effects due to absorption. We have measured the spatial distribution of temperature changes during near infrared irradiation (789 nm) as a function of laser power, in both excised tissue (chicken meat and skin) and in the forearm of an awake human volunteer. Light was applied using a 1 mm optical fiber which is characteristic of the topographic system. The temperature of excised chicken tissue increased linearly with power level as 0.097 and 0.042 degreesC/mW at depths of 0 and 1 mm, respectively. Human forearm studies yielded temperature changes of 0.101, 0.038, and 0.030 degreesC/mW at depths of 0.5, 1.0, and 1.5 mm, respectively. Due to direct irradiation of the thermocouple all measurements represent the maximum temperature increase from the laser. In all cases the estimated heating effects from continuous wave optical topography systems were small and well below levels which would endanger tissue cells. The close similarity between ex vivo and in vivo measurements suggests negligible contributions from blood flow in the skin which was further supported by measurements during cuff ischemia. Heating effects decreased sharply with both depth and lateral position; thus, for optode spacings greater than a few millimeters, fibers can be treated independently. Finite element analysis confirms that the experimental results are consistent with a simple heat conduction model. (C) 2000 Society of Photo-Optical Instrumentation Engineers. [S1083-3668(00)00104-0].
引用
收藏
页码:383 / 390
页数:8
相关论文
共 33 条
[1]   EFFECTS OF 830 NM CONTINUOUS-WAVE LASER-DIODE IRRADIATION ON MEDIAN NERVE FUNCTION IN NORMAL SUBJECTS [J].
BASFORD, JR ;
HALLMAN, HO ;
MATSUMOTO, JY ;
MOYER, SK ;
BUSS, JM ;
BAXTER, GD .
LASERS IN SURGERY AND MEDICINE, 1993, 13 (06) :597-604
[2]  
BASFORD JR, 1990, LASER SURG MED, V10, P25
[3]  
BIRNGRUBER R, 1984, T OPHTHAL SOC UK, V103, P422
[4]   COGNITION-ACTIVATED LOW-FREQUENCY MODULATION OF LIGHT-ABSORPTION IN HUMAN BRAIN [J].
CHANCE, B ;
ZHUANG, Z ;
UNAH, C ;
ALTER, C ;
LIPTON, L .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1993, 90 (08) :3770-3774
[5]   SYSTEM FOR LONG-TERM MEASUREMENT OF CEREBRAL BLOOD AND TISSUE OXYGENATION ON NEWBORN-INFANTS BY NEAR-INFRARED TRANS-ILLUMINATION [J].
COPE, M ;
DELPY, DT .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (03) :289-294
[6]   Effects of near-infrared radiation on the epidermal proliferation and cutaneous immune function in mice [J].
Danno, K ;
Sugie, N .
PHOTODERMATOLOGY PHOTOIMMUNOLOGY & PHOTOMEDICINE, 1996, 12 (06) :233-236
[7]   CELLULAR RESPONSES TO COMBINATIONS OF HYPERTHERMIA AND RADIATION [J].
DEWEY, WC ;
HOPWOOD, LE ;
SAPARETO, SA ;
GERWECK, LE .
RADIOLOGY, 1977, 123 (02) :463-474
[8]  
DEWEY WC, 1977, P 2 INT S ESS GERM J
[9]  
Dusinberre G.M., 1961, Heat Transfer Calculations by Finite Differences
[10]  
FINSEN N, 1899, BEDEUTUNG CHEM STRAH