Immunodetection by quartz crystal microbalance -: A new approach for direct detection of rabbit IgG and peroxidase

被引:58
作者
Bizet, K [1 ]
Gabrielli, C [1 ]
Perrot, H [1 ]
机构
[1] Univ Paris 06, UPR 15 CNRS, F-75252 Paris 05, France
关键词
quartz crystal microbalance; peroxidase; rabbit IgG; immunodetection; 27-MHz resonators;
D O I
10.1385/ABAB:89:2-3:139
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Biodetection is one of the most important challenges for the twenty-first century: many fields are concerned, mainly environmental and medical. The quartz crystal microbalance (QCM) may of for great possibilities for this purpose: a direct response signal, which characterizes the binding event between a sensitive layer, immobilized onto the surface transducer, and the analyte to be detected, can be obtained. However, for the detection of small biomolecules such as antigens, it is quite difficult to obtain an observable signal that corresponds directly to the binding event. In general, this is owing to the lack of mass sensitivity of the commonly used QCM, with 5- to 10-MHz quartz crystals. For improving this mass sensitivity, a 27-MHz quartz resonator was developed and incorporated in a flow-through microcell. Two biospecies, IgG rabbit and peroxidase enzyme, were studied with this ultrasensitive QCM in terms of specificity, detection limit, and calibration curve.
引用
收藏
页码:139 / 149
页数:11
相关论文
共 12 条
[1]   STUDY OF IMMUNOGLOBULIN-G THIN-LAYERS OBTAINED BY THE LANGMUIR-BLODGETT METHOD - APPLICATION TO IMMUNOSENSORS [J].
BARRAUD, A ;
PERROT, H ;
BILLARD, V ;
MARTELET, C ;
THERASSE, J .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (01) :39-48
[2]   Piezoelectric immunosensor for the detection of immunoglobulin M [J].
Chu, X ;
Lin, ZH ;
Shen, GL ;
Yu, RQ .
ANALYST, 1995, 120 (12) :2829-2832
[3]   CALIBRATION OF THE ELECTROCHEMICAL QUARTZ CRYSTAL MICROBALANCE [J].
GABRIELLI, C ;
KEDDAM, M ;
TORRESI, R .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1991, 138 (09) :2657-2660
[4]   A PIEZOELECTRIC IMMUNOBIOSENSOR FOR ATRAZINE IN DRINKING-WATER [J].
GUILBAULT, GG ;
HOCK, B ;
SCHMID, R .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (06) :411-419
[5]  
HARTEVELD JL, 1997, BIOSENS BIOELECTRON, V2, P661
[6]   DEVELOPMENT OF A PIEZOELECTRIC IMMUNOSENSOR FOR THE DETECTION OF HUMAN ERYTHROCYTES [J].
KONIG, B ;
GRATZEL, M .
ANALYTICA CHIMICA ACTA, 1993, 276 (02) :329-333
[7]   A novel method of immobilizing antibodies on a quartz crystal microbalance using plasma-polymerized films for immunosensors [J].
Nakanishi, K ;
Muguruma, H ;
Karube, I .
ANALYTICAL CHEMISTRY, 1996, 68 (10) :1695-1700
[8]   INTEGRATED OPTICAL INPUT GRATING COUPLERS AS BIOCHEMICAL SENSORS [J].
NELLEN, PM ;
TIEFENTHALER, K ;
LUKOSZ, W .
SENSORS AND ACTUATORS, 1988, 15 (03) :285-295
[9]   Evaluation of binding of human serum albumin (HSA) to monoclonal and polyclonal antibody by means of piezoelectric immunosensing technique [J].
Sakai, G ;
Saiki, T ;
Uda, T ;
Miura, N ;
Yamazoe, N .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 42 (02) :89-94
[10]   SELECTIVE AND REPEATABLE DETECTION OF HUMAN SERUM-ALBUMIN BY USING PIEZOELECTRIC IMMUNOSENSOR [J].
SAKAI, G ;
SAIKI, T ;
UDA, T ;
MIURA, N ;
YAMAZOE, N .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 24 (1-3) :134-137