A sheathless nanoflow electrospray interface for on-line capillary electrophoresis mass spectrometry

被引:77
作者
Janini, GM [1 ]
Conrads, TP [1 ]
Wilkens, KL [1 ]
Issaq, HJ [1 ]
Veenstra, TD [1 ]
机构
[1] NCI, Analyt Chem Lab, SAIC Frederick Inc, Frederick, MD 21702 USA
关键词
D O I
10.1021/ac020661+
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel, rugged capillary electrophoresis-electrospray ionization (CE-ESI) interface where the separation column, an electrical porous junction, and the spray tip are integrated on a single piece of a fused-silica capillary is described. ESI is accomplished by applying an electrical potential through an easily prepared porous junction across a 3-4-mm length of fused silica. A stable electrospray is produced at nanoflow rates generated in the capillary by electrophoretic and electroosmotic forces. The interface is particularly well suited for the detection of low-femtomole levels of proteins and peptides. The ruggedness of this interface was evident by the continuous operation of the same column for over a 2-week period with no detectable deterioration in separation or electrospray performance. The new interface was used for the LC-ESI-MS separation and analysis of peptides and proteins. Injection of 25 fmol of [Glu(1)]-fibrinopeptide B using the new device produced a CE-ESI-MS electropherogram with a signal-to-noise ratio of over 100 for this peptide.
引用
收藏
页码:1615 / 1619
页数:5
相关论文
共 32 条
[1]  
Barroso MB, 1999, J AM SOC MASS SPECTR, V10, P1271
[2]   Analysis of peptides, proteins, protein digests, and whole human blood by capillary electrophoresis electrospray ionization mass spectrometry using an in-capillary electrode sheathless interface [J].
Cao, P ;
Moini, M .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1998, 9 (10) :1081-1088
[3]   Sheathless capillary electrophoresis/electrospray mass spectrometry using a carbon-coated fused silica capillary [J].
Chang, YZ ;
Her, GR .
ANALYTICAL CHEMISTRY, 2000, 72 (03) :626-630
[4]   Advances in CE MS [J].
Ding, JM ;
Vouros, P .
ANALYTICAL CHEMISTRY, 1999, 71 (11) :378A-385A
[5]   ONLINE TIME-OF-FLIGHT MASS-SPECTROMETRIC ANALYSIS OF PEPTIDES SEPARATED BY CAPILLARY ELECTROPHORESIS [J].
FANG, LL ;
ZHANG, R ;
WILLIAMS, ER ;
ZARE, RN .
ANALYTICAL CHEMISTRY, 1994, 66 (21) :3696-3701
[6]   Protein identification by capillary zone electrophoresis/microelectrospray ionization-tandem mass spectrometry at the subfemtomole level [J].
Figeys, D ;
vanOostveen, I ;
Ducret, A ;
Aebersold, R .
ANALYTICAL CHEMISTRY, 1996, 68 (11) :1822-1828
[7]   Protein identification by solid phase microextraction-capillary zone electrophoresis-microelectrospray-tandem mass spectrometry [J].
Figeys, D ;
Ducret, A ;
Yates, JR ;
Aebersold, R .
NATURE BIOTECHNOLOGY, 1996, 14 (11) :1579-1583
[8]   Interfaces for coupled liquid-phase separation/mass spectrometry techniques.: An update on recent developments [J].
Gelpí, E .
JOURNAL OF MASS SPECTROMETRY, 2002, 37 (03) :241-253
[9]   Amperometric detection in capillary electrophoresis with on etched joint [J].
Hu, S ;
Wang, ZL ;
Li, PB ;
Cheng, JK .
ANALYTICAL CHEMISTRY, 1997, 69 (02) :264-267
[10]   Capillary zone electrophoresis-electrospray mass spectrometry at submicroliter flow rates: Practical considerations and analytical performance [J].
Kelly, JF ;
Ramaley, L ;
Thibault, P .
ANALYTICAL CHEMISTRY, 1997, 69 (01) :51-60