Determination of organic peroxides by liquid chromatography with on-line post-column ultraviolet irradiation and peroxyoxalate chemiluminescence detection

被引:45
作者
Wada, M
Inoue, K
Ihara, A
Kishikawa, N
Nakashima, K
Kuroda, N
机构
[1] Nagasaki Univ, Course Pharmaceut Sci, Grad Sch Biomed Sci, Nagasaki 8528521, Japan
[2] Akita Univ, Fac Educ & Human Studies, Akita 0108502, Japan
关键词
peroxyoxalate chemiluminescence detection; organic peroxides; peroxides;
D O I
10.1016/S0021-9673(02)01473-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A HPLC method was developed for the simultaneous determination of organic peroxides and hydrogen peroxide with peroxyoxalate chemiluminescence (PO-CL) detection following on-line UV irradiation. Organic peroxides [i.e., benzoyl peroxide (BP), tert-butyl hydroperoxide (BHP), tert.-butyl perbenzoate (BPB), cumene hydroperoxide (CHP)] were UV irradiated (254 nm, 15 W) to generate hydrogen peroxide, which was determined by PO-CL detection. The conditions for UV irradiation and PO-CL detection were optimized by a flow injection analysis (FIA) system. Generation of hydrogen peroxide from peroxides with on-line UV irradiation also was confirmed by the FIA system by incorporating an enzyme column reactor immobilized with catalase. The separation of four organic peroxides and hydrogen peroxide by HPLC was accomplished isocratically on an ODS column within 30 min. The detection limits (signal-to-noise ratio=3) were 1.1 muM for hydrogen peroxide, 6.8 muM for BP, 31.3 muM for BHP, 7.5 muM for BPB and 1.3 muM for CHP. The proposed method was applied to the determination of BP in wheat flour. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:189 / 195
页数:7
相关论文
共 20 条
[1]   SIMULTANEOUS DETERMINATION OF HYDROPEROXIDES OF PHOSPHATIDYLCHOLINE, CHOLESTEROL ESTERS AND TRIACYLGLYCEROLS BY COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH A POSTCOLUMN DETECTION SYSTEM [J].
AKASAKA, K ;
OHRUI, H ;
MEGURO, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 622 (02) :153-159
[2]   DIETARY CARCINOGENS AND ANTICARCINOGENS - OXYGEN RADICALS AND DEGENERATIVE DISEASES [J].
AMES, BN .
SCIENCE, 1983, 221 (4617) :1256-1264
[3]   Determination of benzoyl peroxide in wheat flour by ion chromatography with precolumn derivatization [J].
Chen, QC ;
Mou, SF ;
Hou, XP ;
Ni, ZM .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 1998, 21 (05) :705-716
[4]   Determination of hydrogen peroxide in workplace air: interferences and method validation [J].
Christensen, CS ;
Brodsgaard, S ;
Mortensen, P ;
Egmose, K ;
Linde, SA .
JOURNAL OF ENVIRONMENTAL MONITORING, 2000, 2 (04) :339-343
[5]   QUANTITATIVE-DETERMINATION OF BENZOYL PEROXIDE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND COMPARISON TO THE IODOMETRIC METHOD [J].
GADDIPATI, N ;
VOLPE, F ;
ANTHONY, G .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1983, 72 (12) :1398-1400
[6]   Determination of H2O2 and organic peroxides by high-performance liquid chromatography with post-column UV irradiation, derivatization and fluorescence detection [J].
Hong, JG ;
Maguhn, J ;
Freitag, D ;
Kettrup, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1998, 361 (02) :124-128
[7]   Detection of volatile organic peroxides in indoor air [J].
Hong, JG ;
Maguhn, J ;
Freitag, D ;
Kettrup, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 371 (07) :961-965
[8]   Automated measurement system for H2O2 in the atmosphere by diffusion scrubber sampling and HPLC analysis of Ti(IV)-PAR-H2O2 complex [J].
Komazaki, Y ;
Inoue, T ;
Tanaka, S .
ANALYST, 2001, 126 (05) :587-593
[9]  
Korytowski W, 1999, METHOD ENZYMOL, V300, P23
[10]  
MIYAZAWA T, 1994, METHOD ENZYMOL, V233, P324