Interfacial kinetics of titanium- and cobalt-based implant alloys in human serum: Metal release and biofilm formation

被引:53
作者
Hallab, NJ [1 ]
Skipor, A [1 ]
Jacobs, JJ [1 ]
机构
[1] Rush Presbyterian St Lukes Med Ctr, Dept Orthoped Surg, Chicago, IL 60612 USA
关键词
biofilm; metal release; corrosion; metal-protein binding; interfacial kinetics;
D O I
10.1002/jbm.a.10429
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The biocompatibility of metallic implant surfaces is governed in large part by the interfacial kinetics associated with metal release and protein binding. The kinetics of metal release from, and protein binding to, cobalt- and titanium-based implant alloys in human serum were investigated by (1) measuring the temporal release of Cr and Ti into serum from Co-Cr-Mo (ASTM F-75) and Ti implant alloys (Ti-6Al-4V: ASTM F136, and commercially pure Ti, cpTi: ASTM F67), respectively, (2) examining the composition of human serum proteins adsorbed onto the surfaces of Co- and Ti-based implant alloys; and (3) identifying the serum proteins associated with the binding of soluble Cr and Ti degradation products. Analysis of metal dissolution kinetics found that Cr was released from Co-based implant alloy at an order of magnitude higher than Ti was released from Ti-based implant alloys. Serum became saturated with soluble CR and Ti at levels as high as 3,250 ng/mL Ti from cpTi; 3,7550 ng/mL Ti from Ti-6Al-4V; and 35,400 ng/mL Cr from Co-Cr-Mo degradation. The observation that human serum binds more released metal from Co-based alloy dissolution was consistent with the observed differences in bio-film composition between the two alloys, where additional serum protein(s) of approximately approximate to140 (kDa) molecular weight were detected on Co-based implant alloy surfaces. However, both Cr and Ti released from Co- and Ti-based alloys exhibited a bimodal binding pattern to both low molecular weight serum protein(s) (<32 kDa), and to higher molecular weight protein(s) in the 180-250 kDa range. Identification of metal alloy-dependent biofilm compositions and dissolution products provides the basis for understanding the bioavailability and bioreactivity of these implant alloys and their degradation products. (C) 2003 Wiley Periodicals, Inc. J Biomed Mater Res 65A: 311-318, 2003.
引用
收藏
页码:311 / 318
页数:8
相关论文
共 48 条
[1]  
Andrade J., 1986, Protein adsorption and materials biocompatibility: A tutorial review and suggested hypotheses
[2]  
[Anonymous], ORTHOPAEDIC BIOMATER
[3]   SYSTEMIC EFFECTS OF BIOMATERIALS [J].
BLACK, J .
BIOMATERIALS, 1984, 5 (01) :11-18
[4]  
DORR LD, 1990, CLIN ORTHOP RELAT R, V261, P82
[5]  
Edwards DL, 1998, J ORAL REHABIL, V25, P2
[6]   COMPETITION BETWEEN ADSORBED FIBRINOGEN AND HIGH-MOLECULAR-WEIGHT KININOGEN ON SOLID-SURFACES INCUBATED IN HUMAN-PLASMA (THE VROMAN EFFECT) - INFLUENCE OF SOLID-SURFACE WETTABILITY [J].
ELWING, H ;
ASKENDAL, A ;
LUNDSTROM, I .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1987, 21 (08) :1023-1028
[7]   PROTEIN ADSORPTION ON LOW-TEMPERATURE ISOTROPIC CARBON .1. PROTEIN CONFORMATIONAL CHANGE PROBED BY DIFFERENTIAL SCANNING CALORIMETRY [J].
FENG, L ;
ANDRADE, JD .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (06) :735-743
[8]  
Hallab NJ, 2000, J BIOMED MATER RES, V49, P353, DOI 10.1002/(SICI)1097-4636(20000305)49:3<353::AID-JBM8>3.0.CO
[9]  
2-T
[10]  
Hallab NJ, 2001, J BIOMED MATER RES, V56, P427, DOI 10.1002/1097-4636(20010905)56:3<427::AID-JBM1112>3.3.CO