Simultaneous determination of aflatoxins and ochratoxin A in food using a fully automated immunoaffinity column clean-up and liquid chromatography-fluorescence detection

被引:86
作者
Chan, D [1 ]
MacDonald, SJ [1 ]
Boughtflower, V [1 ]
Brereton, P [1 ]
机构
[1] Cent Sci Lab, York YO41 1LZ, N Yorkshire, England
关键词
food analysis; cereals; nuts; mycotoxins; aflatoxins; ochratoxin A;
D O I
10.1016/j.chroma.2004.09.096
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An automated HPLC method for the simultaneous detection of aflatoxins (AF) and ochratoxin A (OA) has been developed. The method uses an immunoaffinity column containing antibodies specific to both AF and OA. The samples were extracted with an acetonitrile/water mixture and diluted with phosphate buffer saline (PBS). The aqueous extracts were then transferred to an ASPEC HPLC system for automated clean-up using AflaOchra(TM) immunoaffinity columns. OA and AF were quantified using HPLC with fluorescence detection, with a run time of approximately 40 min. Limits of quantification were estimated as 0.2mug/kg for OA and AFB(1), AFB(2), AFG(1) and AFG(2). Initial validation of this method gave average recoveries of 72-101% for OA and AF for a range of food products (maize cereal products and peanut butter). Within laboratory RSDr and RSDR for a 5.0 mug/kg spike level in maize cereals was found to be 7.6-10.1% (AF and OA) and 10.2-13.8%, respectively. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:13 / 16
页数:4
相关论文
共 12 条