Different efficacy of various blocking reagents to eliminate interferences by human antimouse antibodies with a two-site immunoassay

被引:35
作者
Reinsberg, J
机构
[1] Zentrum Frauenheilkunde Geburtshilfe, Universität Bonn, D-53127 Bonn
关键词
CA-125; immunoassay; interference; human antimouse antibodies;
D O I
10.1016/0009-9120(95)02044-6
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: The efficacy of three reagents to eliminate interferences with the cancer antigen 125 (CA-125) assay Enzymun CA-125 II by human antimouse antibodies (HAMA) formed by patients after injection of the murine anti-CA-125 antibody OC125 is compared. Methods: Apparent CA-125 concentrations of 14 serum samples obtained from 6 patients after multiple injections of 1 mg radiolabeled OC125 F(ab')(2) fragments were measured with the Enzymun CA-125 II before and after preincubation, either with nonspecific mouse IgG, with the polymerized mouse IgG MAK-33, or with the commercially available HAMA-blocking reagent IIR. Results: In all samples with HAMA concentrations ranging from 341 to 46900 mu g/L, false-positive CA-125 values were measured with the Enzymun CA-125 II, which could be reduced by preincubation with the HAMA-blocking reagents. However, although after preincubation with 2 g/L IIR for all samples the CA-125 concentrations measured were reduced to values within the normal range, after preincubation with 0.7 g/L of polyclonal mouse IgG for five samples and after preincubation with 0.7 g/L of MAK-33 for all samples also the reduced values were considerably elevated. Larger amounts of mouse IgG or MAK-33]ed only to a slight reduction of the remaining false-positive CA-125 values. Conclusion: The present results demonstrate that the polymerized mouse IgG MAK-33 and also the normal murine IgG are not suitable for completely eliminating interferences by HAMAs formed after OC125 treatment. Only the HAMA-blocking reagent IIR seems to be an effective agent to eliminate these interferences. Thus, further studies of this material as a blocking reagent seem to be warranted.
引用
收藏
页码:145 / 148
页数:4
相关论文
共 16 条
  • [1] BIENHAUS G, 1990, RECENT RESULTS TUMOR, P469
  • [2] BOERMAN OC, 1990, CLIN CHEM, V36, P888
  • [3] BOSCATO LM, 1986, CLIN CHEM, V32, P1491
  • [4] BOSCATO LM, 1988, CLIN CHEM, V34, P27
  • [5] HOWANITZ PJ, 1982, CLIN CHEM, V28, P427
  • [6] KENEMANS P, 1993, CLIN CHEM, V39, P2509
  • [7] KRICKA LJ, 1990, CLIN CHEM, V36, P892
  • [8] INTERFERENCE BY HUMAN-ANTIBODIES WITH TUMOR-MARKER ASSAYS
    REINSBERG, J
    [J]. HYBRIDOMA, 1995, 14 (02): : 205 - 208
  • [9] REINSBERG J, 1994, CLIN CHEM, V40, P951
  • [10] REINSBERG J, 1993, CLIN CHEM, V39, P891