Measurement of the performance of nerve cuff electrodes for recording

被引:21
作者
Andreasen, LNS [1 ]
Struijk, JJ
Lawrence, S
机构
[1] Univ Aalborg, Ctr Sensory Motor Interact, Aalborg, Denmark
[2] Univ Utah, Dept Biomed Engn, Salt Lake City, UT 84112 USA
关键词
cuff electrode; nerve fibre; action potential; model;
D O I
10.1007/BF02345015
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
New designs of cuff electrodes for the recording of signals from peripheral nerves are typically tested in acute animal experiments before long-term evaluation takes place. A reproducible, cost-effective and fast method is presented for evaluating cuff electrodes with respect to signal amplitude, noise rejection, and, in some cases, selectivity, as an alternative to acute in vivo experiments. Comparisons with a computer model and with signals obtained from rabbit tibial nerve give good agreement with the new method. It is shown that an imperfect closure of the cuff around the nerve can easily lead to more than 50% loss of the signal amplitude. Noise from sources external to the cuff is not significantly affected by the closing mechanism, but is strongly reduced by a tripolar cuff configuration as compared with a monopolar one (reduction factor 2.8 to 58, mean = 6.5, n = 6). in dual-channel cuffs, cross-talk is below 1.2% indicating a very high selectivity.
引用
收藏
页码:447 / 453
页数:7
相关论文
共 20 条
[1]  
CHIU SY, 1979, J PHYSIOL-LONDON, V292, P149, DOI 10.1113/jphysiol.1979.sp012843
[2]   COMPOUND ACTION POTENTIALS RECORDED FROM MAMMALIAN PERIPHERAL-NERVES FOLLOWING LIGATION OR RE-SUTURING [J].
DAVIS, LA ;
GORDON, T ;
HOFFER, JA ;
JHAMANDAS, J ;
STEIN, RB .
JOURNAL OF PHYSIOLOGY-LONDON, 1978, 285 (DEC) :543-559
[3]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[4]   Restoration of lateral hand grasp using natural sensors [J].
Haugland, M ;
Lickel, A ;
Riso, R ;
Adamczyk, MM ;
Keith, M ;
Jensen, IL ;
Haase, J ;
Sinkjaer, T .
ARTIFICIAL ORGANS, 1997, 21 (03) :250-253
[5]  
Haugland M., 1996, 18 ANN INT C IEEE EN
[6]  
Haugland M. K., 1995, IEEE Transactions on Rehabilitation Engineering, V3, P307, DOI 10.1109/86.481970
[7]  
Hoffer J A., 1990, Neurophysiological Techniques: Applications to neural systems, Vvol 15, P65
[8]   Analysis of bladder related nerve cuff electrode recordings from preganglionic pelvic nerve and sacral roots in pigs [J].
Jezernik, S ;
Wen, JG ;
Rijkhoff, NJM ;
Djurhuus, JC ;
Sinkjær, T .
JOURNAL OF UROLOGY, 2000, 163 (04) :1309-1314
[9]  
Kallesoe K., 1996, Patent No. [US5487756, 5487756]
[10]   Cuff electrodes for chronic stimulation and recording of peripheral nerve activity [J].
Loeb, GE ;
Peck, RA .
JOURNAL OF NEUROSCIENCE METHODS, 1996, 64 (01) :95-103