Polyindole modified potassium ion sensor using dibenzo-18-crown-6 mediated PVC matrix membrane

被引:78
作者
Pandey, PC [1 ]
Prakash, R [1 ]
机构
[1] Banaras Hindu Univ, Dept Chem, Varanasi 221005, Uttar Pradesh, India
关键词
ion-selective electrode; conducting polymer; polyindole; potentiometry;
D O I
10.1016/S0925-4005(97)00332-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A potassium ion-selective electrode using PVC matrix membrane impregnated with dibenzo-18-crown-6 at the surface of the polyindole modified electrode is reported. The electroactivity of the polyindole film contributing to the high selectivity of the neutral carrier based ion-sensor is analyzed based on the cyclic voltammetry. The polyindole modified electrode is formed by the electropolymerization of the indole monomer in dichloromethane containing tetrabutyl ammonium perchlorate (TBAP). A highly stable electroactive polyindole film is obtained. A neutral carrier impregnated PVC membrane is formed by casting over the surface of the polyindole modified electrode. The resulting ion-selective electrode without the incorporation of polyanions shows high reproducibility with negligible drift of the standard potential with a slope of 53 mV/decade. The lowest detection limit for the potassium ion sensor is 7.0 x 10(-6) mol dm(-3) with a wide linearity over almost five decades. A typical response of the neutral carrier impregnated PVC membrane over the polyindole modified electrode to the potassium ion is reported. (C) 1998 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:61 / 65
页数:5
相关论文
共 15 条
[1]   THE ELECTROCHEMISTRY OF CYTOCHROME-C AT A CONDUCTING POLYMER ELECTRODE [J].
BARTLETT, PN ;
FARINGTON, J .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 261 (2B) :471-475
[2]   ALL-SOLID-STATE SODIUM-SELECTIVE ELECTRODE BASED ON A CALIXARENE IONOPHORE IN A POLY(VINYL CHLORIDE) MEMBRANE WITH A POLYPYRROLE SOLID CONTACT [J].
CADOGAN, A ;
GAO, ZQ ;
LEWENSTAM, A ;
IVASKA, A ;
DIAMOND, D .
ANALYTICAL CHEMISTRY, 1992, 64 (21) :2496-2501
[3]  
CAMPANELLA L, 1986, CHIM IND-MILAN, V68, P71
[4]  
CAMPANELLA L, 1988, ANALUSIS S9, V16, P120
[5]  
*COMM AN NOM, 1975, PURE APPL CHEM, V48, P129
[6]   ION-SELECTIVE ELECTRODES USING AN IONOPHORE COVALENTLY ATTACHED TO CARBOXYLATED POLY(VINYL CHLORIDE) [J].
DAUNERT, S ;
BACHAS, LG .
ANALYTICAL CHEMISTRY, 1990, 62 (14) :1428-1431
[7]   SPECTROELECTROCHEMICAL AND EPR DETERMINATION OF THE NUMBER OF ELECTRONS TRANSFERRED IN REDOX PROCESSES IN ELECTROACTIVE POLYMERS - POLYINDOLE FILMS [J].
JACKOWSKA, K ;
KUDELSKI, A ;
BUKOWSKA, J .
ELECTROCHIMICA ACTA, 1994, 39 (10) :1365-1368
[8]   POTENTIOMETRIC DIGOXIN ANTIBODY MEASUREMENTS WITH ANTIGEN IONOPHORE BASED MEMBRANE ELECTRODES [J].
KEATING, MY ;
RECHNITZ, GA .
ANALYTICAL CHEMISTRY, 1984, 56 (04) :801-806
[9]   ELECTROCHEMICAL CYCLING AND ELECTROCHROMIC PROPERTIES OF POLYINDOLE [J].
MAAROUF, EB ;
BILLAUD, D ;
HANNECART, E .
MATERIALS RESEARCH BULLETIN, 1994, 29 (06) :637-643
[10]  
MOMMA T, 1995, SENSOR ACTUAT B-CHEM, V24, P724