Variability and repeatability of retinal blood flow measurements using the Canon Laser Blood Flowmeter

被引:51
作者
Guan, K
Hudson, C [1 ]
Flanagan, JG
机构
[1] Univ Toronto, Dept Ophthalmol, Toronto, ON M5T 2S8, Canada
[2] Univ Waterloo, Sch Optometry, Waterloo, ON N2L 3G1, Canada
关键词
D O I
10.1016/S0026-2862(03)00007-4
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The purpose of this work was to determine the within-session variability and between-session repeatability of the Canon Laser Blood Flowmeter (CLBF), Model 100, an instrument that permits the noninvasive measurement of retinal blood flow. The CLBF calculates flow in mul/min based on the Poiseuille principle. One eye of 20 normal subjects (mean age 36.5; SD 9.7 years) was randomly chosen. A minimum of five measurements was acquired of a temporal arteriole approximately 1 disc diameter from the optic nerve. Measurements were repeated within a 1-month period. Blood pressure and intraocular pressure were measured. The coefficient of variation (COV) and the coefficient of repeatability (COR) were calculated for each individual. The individual COVs for diameter, velocity, and flow ranged from 0.5 to 6.5% (median 2.0%), 4.8 to 39.7% (median 19.9%), and 4.8 to 37.3% (median 19.3%), respectively. The group mean CORs for diameter, velocity, and flow were 5.2 mum (relative to a mean effect of 104.6 mum), 8.8 mm/s (relative to a mean effect of 33.9 mm/s), and 2.6 mul/min (relative to a mean effect of 8.8 mul/min), respectively. The CLBF gave consistent and repeatable measurements of blood flow within retinal arterioles in normal subjects. Given the range of individual variation in the velocity measurement, and thus flow, confidence limits for retinal hemodynamics need to be determined on an individual basis. (C) 2003 Elsevier Science (USA). All rights reserved.
引用
收藏
页码:145 / 151
页数:7
相关论文
共 24 条
[1]  
Alm A., 1992, Adler's physiology of the eye, P198
[2]   Bowman Lecture 1998 - Diabetic retinopathy: some cellular, molecular and therapeutic considerations [J].
Archer, DB .
EYE, 1999, 13 (4) :497-523
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Bursell SE, 1996, INVEST OPHTH VIS SCI, V37, P886
[5]  
CANON, 1997, LASER BLOOD FLOWMETE
[6]   EVALUATION OF MICROMETRIC AND MICRODENSITOMETRIC METHODS FOR MEASURING THE WIDTH OF RETINAL VESSEL IMAGES ON FUNDUS PHOTOGRAPHS [J].
DELORI, FC ;
FITCH, KA ;
FEKE, GT ;
DEUPREE, DM ;
WEITER, JJ .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1988, 226 (04) :393-399
[7]   LASER DOPPLER TECHNIQUE FOR ABSOLUTE MEASUREMENT OF BLOOD SPEED IN RETINAL-VESSELS [J].
FEKE, GT ;
GOGER, DG ;
TAGAWA, H ;
DELORI, FC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (09) :673-680
[8]   LASER DOPPLER MEASUREMENTS OF BLOOD VELOCITY IN HUMAN RETINAL-VESSELS [J].
FEKE, GT ;
RIVA, CE .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1978, 68 (04) :526-531
[9]  
FEKE GT, 1985, OPHTHALMOLOGY, V92, P1517
[10]  
FEKE GT, 1989, INVEST OPHTH VIS SCI, V30, P58