Characteristics of a dedicated linear accelerator-based stereotactic radiosurgery radiotherapy unit

被引:54
作者
Das, IJ
Downes, MB
Corn, BW
Curran, WJ
WernerWasik, M
Andrews, DW
机构
[1] WILLS EYE HOSP & RES INST, WILLS NEUROSENSORY INST, DEPT RADIAT ONCOL, PHILADELPHIA, PA 19107 USA
[2] THOMAS JEFFERSON UNIV HOSP, DEPT NEUROSURG, PHILADELPHIA, PA 19107 USA
关键词
stereotactic radiosurgery; stereotactic radiotherapy; linear accelerator; diamond detector; small field dosimetry;
D O I
10.1016/0167-8140(95)01674-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A stereotactic radiosurgery and radiotherapy (SRS/SRT) system on a dedicated Varian Clinac-600SR linear accelerator with Brown-Reberts-Wells and Gill-Thomas-Cosman relocatable Frames along with the Radionics (RSA) planning system is evaluated. The Clinac-600SR has a single 6-MV beam with the same beam characteristics as that of the mother unit, the Clinac-600C. The primary collimator is a fixed cone projecting to a 10-cm diameter at isocenter. The secondary collimator is a heavily shielded cylindrical collimator attached to the face plate of the primary collimator. The tertiary collimation consists of the actual treatment cones. The cone sizes vary from 12.5 to 40.0 mm diameter. The mechanical stability of the entire system was verified. The variations in isocenter position with table, gantry, and collimator rotation were found to be <0.5 mm with a compounded accuracy of less than or equal to 1.0 mm. The radiation leakage under the cones was < 1(a)` o measured at a depth of 5 cm in a phantom. The beam profiles of all cones in the x and y directions were within +/- 0.5 mm and match with the physical size of the cone. The dosimetric data such as tissue maximum ratio, off-axis ratio, and cone factor were taken using film, diamond detector, and ion chambers. The mechanical and dosimetric characteristics including dose linearity of this unit are presented and found to be suitable for SRS/SRT. The difficult in absolute dose measurement For small cone is discussed.
引用
收藏
页码:61 / 68
页数:8
相关论文
共 31 条
  • [1] [Anonymous], 1983, MED PHYS, V10, P741
  • [2] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) : 1901 - 1913
  • [3] A SURVEY OF METHODS TO CALCULATE MONITOR SETTINGS
    BJARNGARD, BE
    BARDEROMA, R
    CORRAO, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03): : 749 - 752
  • [4] FRACTIONATED REGIMENS FOR STEREOTAXIC RADIOTHERAPY OF RECURRENT TUMORS IN THE BRAIN
    BRENNER, DJ
    MARTEL, MK
    HALL, EJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03): : 819 - 824
  • [5] HALO RING SUPPORTING THE BROWN-ROBERTS-WELLS STEREOTAXIC FRAME FOR FRACTIONATED RADIOTHERAPY
    CARINI, S
    SCIELZO, G
    RUGGIERI, FG
    BISTOLFI, F
    RAVEGNANI, M
    ANDREUSSI, L
    [J]. ACTA NEUROCHIRURGICA, 1994, 129 (1-2) : 92 - 96
  • [6] A HALO-RING TECHNIQUE FOR FRACTIONATED STEREOTAXIC RADIOTHERAPY
    CLARK, BG
    PODGORSAK, EB
    SOUHAMI, L
    OLIVIER, A
    SIXEL, KE
    CARON, JL
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (786) : 522 - 527
  • [7] THE DEPENDENCE OF MASS-ENERGY ABSORPTION-COEFFICIENT RATIOS ON BEAM SIZE AND DEPTH IN A PHANTOM
    CUNNINGHAM, JR
    WOO, M
    ROGERS, DWO
    BIELAJEW, AF
    [J]. MEDICAL PHYSICS, 1986, 13 (04) : 496 - 502
  • [8] DOSIMETRIC ACCURACY AT LOW MONITOR UNIT SETTINGS
    DAS, IJ
    KASE, KR
    TELLO, VM
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (765) : 808 - 811
  • [9] DAS IJ, 1995, UNPUB PHYS MED BIOL
  • [10] FRACTIONATED RADIOTHERAPY OF SMALL INOPERABLE LESIONS OF THE BRAIN USING A NONINVASIVE STEREOTAXIC FRAME
    DELANNES, M
    DALY, NJ
    BONNET, J
    SABATIER, J
    TREMOULET, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03): : 749 - 755