Response of a new ceramic-oxynitride (Cernox) resistance temperature sensor in high magnetic fields

被引:13
作者
Ramsbottom, HD
Ali, S
Hampshire, DP
机构
[1] Superconductivity Group, Department of Physics, University of Durham, Durham DH1 3LE, South Road
关键词
thermometry; high magnetic fields;
D O I
10.1016/0011-2275(96)80772-7
中图分类号
O414.1 [热力学];
学科分类号
摘要
The resistance of a commercial Cernox (ceramic-oxynitride) thermometer has been measured from 4.2 to 30 K in magnetic fields up to 15 T and compared to that of a carbon-glass thermometer. Throughout this field and temperature range, the maximum magnetic field-induced change in the zero-field calibration of the Cernox thermometer is less than 200 mK. This is a factor of two lower than that of the carbon-glass thermometer. Furthermore, its sensitivity at room temperature is a factor of two higher.
引用
收藏
页码:61 / 63
页数:3