Analysis of pedicle and translaminar facet fixation in a multisegment interbody fusion model

被引:35
作者
Eskander, Mark
Brooks, Dahari
Ordway, Nat
Dale, Elizabeth
Connolly, Patrick
机构
[1] Univ Massachusetts, Mem Med Ctr, Arthrit & Total Joint Replacement Ctr, Worcester, MA 01605 USA
[2] SUNY Upstate Med Univ, Dept Orthoped Surg, Syracuse, NY USA
[3] Columbia Univ, Coll Phys & Surg, New York, NY 10027 USA
关键词
multisegment interbody fusion; translaminar screws; pedicle screws; biomechanics;
D O I
10.1097/01.brs.0000259232.30302.bf
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. This is a biomechanical study. We compared pedicle screws and translaminar screws in the setting of multisegment interbody fusions. Objectives. We investigated the significance of the middle segment pedicle screws in a 2-level interbody fusion model and examined translaminar screw fixation as an alternative to pedicle screw fixation in this 2-level model. Summary of Background Data. Surgical treatment of disc disease focuses on restoration of normal disc height and restriction of abnormal motion. Interbody fusion significantly reduces motion and restores disc space height. Combined anterior and posterior fusion improves fusion rates. Methods. Human cadavers were tested for range of motion ( ROM) and stiffness. Each specimen underwent a 2-level interbody fusion and posterior fixation. There were 3 types of posterior fixation: pedicle screws at 3 levels, pedicle screws at 2 levels, and translaminar facet screws. Biomechanical testing was repeated for each group. Results. The overall ROM decreased after the interbody fusion (P < 0.05). There were no significant differences in construct stiffness for torsion, but there were significant differences when comparing intact to the 3 groups for both flexion and extension (P < 0.05). ROM at L2 and L4 was significantly less for all groups in comparison to the intact condition (P < 0.05). Conclusions. This study supports the omission of the middle segment screws in a 2-level interbody fusion. In addition, translaminar screws are similar to pedicle fixation in a 2-level interbody lumbar fusion.
引用
收藏
页码:E230 / E235
页数:6
相关论文
共 15 条
[1]   Overview of the biology of lumbar spine fusion and principles for selecting a bone graft substitute [J].
Boden, SD .
SPINE, 2002, 27 (16) :S26-S31
[2]   THE ROLE OF FUSION AND INSTRUMENTATION IN THE TREATMENT OF DEGENERATIVE SPONDYLOLISTHESIS WITH SPINAL STENOSIS [J].
BRIDWELL, KH ;
SEDGEWICK, TA ;
OBRIEN, MF ;
LENKE, LG ;
BALDUS, C .
JOURNAL OF SPINAL DISORDERS, 1993, 6 (06) :461-472
[3]   SURGICAL TREATMENT OF SYMPTOMATIC UNSTABLE CERVICAL-SPINE IN RHEUMATOID-ARTHRITIS [J].
FERLIC, DC ;
CLAYTON, ML ;
LEIDHOLT, JD ;
GAMBLE, WE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1975, A 57 (03) :349-354
[4]   A randomized prospective study of posterolateral lumbar fusion - Outcomes with and without pedicle screw instrumentation [J].
France, JC ;
Yaszemski, MJ ;
Lauerman, WC ;
Cain, JE ;
Glover, JM ;
Lawson, KJ ;
Coe, JD ;
Topper, SM .
SPINE, 1999, 24 (06) :553-560
[5]   DEGENERATIVE LUMBAR SPONDYLOLISTHESIS WITH SPINAL STENOSIS - A PROSPECTIVE-STUDY COMPARING DECOMPRESSION WITH DECOMPRESSION AND INTERTRANSVERSE PROCESS ARTHRODESIS [J].
HERKOWITZ, HN ;
KURZ, LT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (06) :802-808
[6]   Translaminar screw fixation of the lumbar and lumbosacral spine - A 5-year follow-up [J].
Humke, T ;
Grob, D ;
Dvorak, J ;
Messikommer, A .
SPINE, 1998, 23 (10) :1180-1184
[7]   Three-dimensional stabilization provided by the external spinal fixator compared to two internal fixation devices: a biomechanical in vitro flexibility study [J].
Lund, T ;
Nydegger, T ;
Rathonyi, G ;
Nolte, LP ;
Schlenzka, D ;
Oxland, TR .
EUROPEAN SPINE JOURNAL, 2003, 12 (05) :474-479
[8]   THE USE OF PRIMARY INTERNAL-FIXATION IN SPONDYLOLISTHESIS [J].
MCGUIRE, RA ;
AMUNDSON, GM .
SPINE, 1993, 18 (12) :1662-1672
[9]  
Sandhu HS, 2000, CLIN ORTHOP RELAT R, P56
[10]  
STEIDEL CC, 1992, APJS, V80, P90