Accuracy of home blood glucose monitors using forearm blood saimples: FreeStyle versus One Touch Ultra

被引:12
作者
Demers, J
Kane, MP
Bakst, G
Busch, RS
Hamilton, RA
机构
[1] Albany Coll Pharm, Dept Pharm Practice, Albany, NY 12208 USA
[2] Endocrine Grp, Albany, NY USA
关键词
dextrose; diabetes mellitus; home tests;
D O I
10.1093/ajhp/60.11.1130
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The accuracy of two home blood glucose monitors using forearm blood samples was studied. Blood samples were obtained from adults by venipuncture and analyzed for glucose concentration. Within five minutes after venipuncture, peripheral blood glucose measurements were taken in duplicate with the FreeStyle and One Touch Ultra monitors by using the same forearm as used for venipuncture. Accuracy was assessed by comparing the results for each first peripheral stick obtained by each monitor with the laboratory reference value. Monitor precision was evaluated by evaluating the mean difference In glucose values between the first and second peripheral sticks. Blood samples were obtained from 250 subjects, 170 (68%) of whom had diabetes mellitus. A total of 98.8% and 98.4% of the blood glucose readings obtained with FreeStyle and One Touch Ultra, respectively, were clinically acceptable, and 65.2% and 59.2% of the measurements, respectively, were within 10% of the laboratory reference values. The average difference between the first and second glucose measurements was 9.7 mg/dL for FreeStyle 5.2 mg/dL for One Touch Ultra. One peripheral stick was needed to obtain an adequate blood sample in 95.6% of subjects using FreeStyle, compared with 83.2% of subjects using One Touch Ultra. Forearm blood glucose measurements obtained with the FreeStyle and One Touch Ultra devices were similar to laboratory reference values obtained by venipuncture.
引用
收藏
页码:1130 / 1135
页数:6
相关论文
共 17 条
[1]  
*AM DIAB ASS, 2002, DIABETES CARE S1, V25, pS97, DOI DOI 10.2337/DIACARE.25.2007.S97
[2]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[3]  
*CLIN CHEM CLIN TO, 2001, M CLIN CHEM CLIN TOX
[4]   Understanding error grid analysis [J].
Cox, DJ ;
GonderFrederick, LA ;
Kovatchev, BP ;
Julian, DM ;
Clarke, WL .
DIABETES CARE, 1997, 20 (06) :911-912
[5]   Rapid changes in postprandial blood glucose produce concentration differences at finger, forearm, and thigh sampling sites [J].
Ellison, JM ;
Stegmann, JM ;
Colner, SL ;
Michael, RH ;
Sharma, MK ;
Ervin, KR ;
Horwitz, DL .
DIABETES CARE, 2002, 25 (06) :961-964
[6]   ACCURACY OF BLOOD-GLUCOSE MONITORING FOR PATIENTS - WHAT IT IS AND HOW TO ACHIEVE IT [J].
FLEMING, DR .
DIABETES EDUCATOR, 1994, 20 (06) :495-&
[7]   SELF-MONITORING OF BLOOD-GLUCOSE BY ADULTS WITH DIABETES IN THE UNITED-STATES POPULATION [J].
HARRIS, MI ;
COWIE, CC ;
HOWIE, LJ .
DIABETES CARE, 1993, 16 (08) :1116-1123
[8]  
Howanitz PJ, 1996, ARCH PATHOL LAB MED, V120, P333
[9]  
Jones BA, 1996, ARCH PATHOL LAB MED, V120, P339
[10]   Glucose monitoring at the arm - Risky delays of hypoglycemia and hyperglycemia detection [J].
Jungheim, K ;
Koschinsky, T .
DIABETES CARE, 2002, 25 (06) :956-960