Improvement of direct determination of Cu and Mn in seawater by GFAAS and total elimination of the saline matrix with the use of hydrofluoric acid

被引:14
作者
Cabon, JY [1 ]
机构
[1] CNRS, UMR 6521, UBO, F-29285 Brest, France
关键词
graphite furnace; seawater; Cu; Mu; HF precipitation of Ca and Mg;
D O I
10.1016/j.talanta.2004.06.041
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Hydrofluoric acid, added to seawater, can assist in the removal of chloride in the drying step by precipitating fluoride salts, thus suppressing the chloride interference effects induced on the atomization signals of Cu and Mn. By adding HF to seawater before the analysis, MgF2 and CaF2 are precipitated at the bottom of the sampling flask, without precipitating Cu and Mn, and are consequently not introduced into the graphite furnace. Because sodium salts are eliminated at the pretreatment step. the whole seawater matrix is eliminated before the atomization of Cu or Mn. Therefore, the analyzed volume of seawater call be increased by using the multi-injection procedure without degradation of the limit of detection and risks of spectral interferences. The limit of detection obtained for Cu and Mn Lire 0.05 and 0.01 mug L-1, respectively, for a 50 muL analyzed seawater volume. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:402 / 407
页数:6
相关论文
共 23 条
[1]   COPPER SIGNALS FROM SEAWATER MATRICES IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY .2. MODEL FOR PHENOMENA OBSERVED WITH SEA-WATER [J].
CABON, JY ;
LEBIHAN, A .
ANALYTICA CHIMICA ACTA, 1987, 198 :103-111
[2]   Determination of Cd and Pb in seawater by graphite furnace atomic absorption spectrometry with the use of hydrofluoric acid as a chemical modifier [J].
Cabon, JY .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (03) :513-524
[3]   Determination of Cu and Mn in seawater by graphite furnace atomic absorption spectrometry with the use of hydrofluoric acid as a chemical modifier [J].
Cabon, JY .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (05) :939-950
[4]   The determination of Cr, Cu and Mn in seawater with transversely heated graphite furnace atomic absorption spectrometry [J].
Cabon, JY ;
LeBihan, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (13) :1703-1716
[5]   DIRECT DETERMINATION OF MANGANESE IN SEAWATER WITH THE LVOV PLATFORM AND ZEEMAN BACKGROUND CORRECTION IN THE GRAPHITE-FURNACE [J].
CARNRICK, GR ;
SLAVIN, W ;
MANNING, DC .
ANALYTICAL CHEMISTRY, 1981, 53 (12) :1866-1872
[6]   Direct determination of cadmium and copper in seawater using a transversely heated graphite furnace atomic absorption spectrometer with Zeeman-effect background corrector [J].
Chan, MS ;
Huang, SD .
TALANTA, 2000, 51 (02) :373-380
[7]   Direct and simultaneous determination of copper, manganese and molybdenum in seawater with a multi-element electrothermal atomic absorption spectrometer [J].
Chen, CY ;
Danadurai, KSK ;
Huang, SD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (04) :404-408
[8]   Spectral phenomena in graphite furnace AAS [J].
deLoosVollebregt, MTC ;
Vrouwe, EX .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (9-10) :1341-1349
[9]  
Ediger R., 1974, ATOMIC ABS NEWS, V13, P61
[10]   THE POSSIBILITIES AND LIMITATIONS OF ELECTROTHERMAL ATOMIZATION IN ATOMIC-ABSORPTION SPECTROPHOTOMETRY FOR THE DIRECT ANALYSIS OF HEAVY-METALS IN SEA-WATER [J].
HOENIG, M ;
WOLLAST, R .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (05) :399-415