Effective dose to patients from chest examinations with tomosynthesis

被引:86
作者
Bath, Magnus [1 ,2 ]
Svalkvist, Angelica [1 ,2 ]
von Wrangel, Alexa [1 ]
Rismyhr-Olsson, Heidi [1 ]
Cederblad, Ake [1 ]
机构
[1] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, SE-41345 Gothenburg, Sweden
[2] Univ Gothenburg, Dept Radiat Phys, SE-41345 Gothenburg, Sweden
关键词
PULMONARY NODULES;
D O I
10.1093/rpd/ncq092
中图分类号
X [环境科学、安全科学];
学科分类号
083001 [环境科学];
摘要
Chest tomosynthesis, which refers to the principle of collecting low-dose projections of the chest at different angles and using these projections to reconstruct section images of the chest, is an imaging technique recently introduced to health care. The main purpose of the present work was to determine the average effective dose to patients from clinical use of chest tomosynthesis. Exposure data for two chest radiography laboratories with tomosynthesis option (Definium 8000 with VolumeRAD option, GE Healthcare, Chalfont St. Giles, UK) were registered for 20 patients with a weight between 60 and 80 kg (average weight of 70.2 kg). The recorded data were used in the Monte Carlo program PCXMC 2.0 (STUK-Radiation and Nuclear Safety Authority, Helsinki, Finland) to determine the average effective dose for each projection. The effective dose for the chest tomosynthesis examination, including a scout view and the tomosynthesis acquisition, was finally obtained by adding the effective doses from all projections. Using the weighting factors given in ICRP 103, the average effective dose for the examination was found to be 0.13 mSv, whereas the average effective dose for the conventional two-view chest radiography examination was 0.05 mSv. A conversion factor of 0.26 mSv Gy(-1) cm(-2) was found suitable for determining the effective dose from a VolumeRAD chest tomosynthesis examination from the total registered kerma-area product. In conclusion, the effective dose to a standard-sized patient (170 cm/70 kg) from a VolumeRAD chest tomosynthesis examination is similar to 2 % of an average chest CT and only two to three times the effective dose from the conventional two-view chest radiography examination.
引用
收藏
页码:153 / 158
页数:6
相关论文
共 17 条
[1]
[Anonymous], 200820 SSMFS SWED RA
[2]
[Anonymous], 1997, Official Journal of the European Communities, V180, P22
[3]
[Anonymous], 1991, Annals of the ICRP, V21
[4]
Digital tomosynthesis of the chest for lung nodule detection: Interim sensitivity results from an ongoing NIH-sponsored trial [J].
Dobbins, James T. ;
McAdams, H. Page ;
Song, Jae-Woo ;
Li, Christina M. ;
Godfrey, Devon J. ;
DeLong, David M. ;
Paik, Sang-Hyun ;
Martinez-Jimenez, Santiago .
MEDICAL PHYSICS, 2008, 35 (06) :2554-2557
[5]
Tomosynthesis imaging: At a translational crossroads [J].
Dobbins, James T., III .
MEDICAL PHYSICS, 2009, 36 (06) :1956-1967
[6]
Digital x-ray tomosynthesis: current state of the art and clinical potential [J].
Dobbins, JT ;
Godfrey, DJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (19) :R65-R106
[7]
*EUR COMM, 1999, RAD PROT, V109
[8]
*INT COMM RAD PROT, 1991, ANN ICRP, V37
[9]
International Commission on Radiological Protection (ICRP), 1996, ANN ICRP, V26, P1, DOI 10.1016/S0146-6453(00)89192-7
[10]
Effective doses in radiology and diagnostic nuclear medicine: A catalog [J].
Mettler, Fred A., Jr. ;
Huda, Walter ;
Yoshizumi, Terry T. ;
Mahesh, Mahadevappa .
RADIOLOGY, 2008, 248 (01) :254-263