Experimental validation of intact and implanted distal femur finite element models

被引:39
作者
Completo, A.
Fonseca, F.
Simoes, J. A. [1 ]
机构
[1] Univ Aveiro, Dept Engn Mecan, P-3810193 Aveiro, Portugal
[2] Univ Beira Interior, Fac Ciencias Saude, Covilha, Portugal
关键词
distal femur; knee prosthesis; composite femur; finite-element analysis; strain gauge measurements;
D O I
10.1016/j.jbiomech.2006.11.012
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
Four finite element (FE) models of intact and distal femur of knee replacements were validated relative to measured bone strains. FE models of linear tetrahedrons were used. Femoral replacements with cemented stemless, cemented and noncemented femoral sterns of the PFC Sigma Modular Knee System were analyzed. Bone strains were recorded at ten locations on the cortex. The magnitude of the FE bone strains corresponded to the mean measured strains, with an overall agreement of 10%. Linear regression between the FE and mean experimental strains produced slopes between 0.94 and 1.06 and R, values between 0.92 and 0.99. RSME values were less than 12%. The FE models were able to adequately replicate the mechanical behavior of distal femur reconstructions. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2467 / 2476
页数:10
相关论文
共 33 条
[1]
Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading [J].
Cheung, G ;
Zalzal, P ;
Bhandari, M ;
Spelt, JK ;
Papini, M .
MEDICAL ENGINEERING & PHYSICS, 2004, 26 (02) :93-108
[2]
A critical analysis of stress shielding evaluation of hip prostheses [J].
Cristofolini, L .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) :409-483
[3]
In vitro testing of a novel limb salvage prosthesis for the distal femur [J].
Cristofolini, L ;
Bini, S ;
Toni, A .
CLINICAL BIOMECHANICS, 1998, 13 (08) :608-615
[4]
Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[5]
Comparison of uniaxial and triaxial rosette gages for strain measurement in the femur [J].
Cristofolini, L ;
Viceconti, M .
EXPERIMENTAL MECHANICS, 1997, 37 (03) :350-354
[6]
Mechanical validation of whole bone composite tibia models [J].
Cristofolini, L ;
Viceconti, M .
JOURNAL OF BIOMECHANICS, 2000, 33 (03) :279-288
[7]
A MEASUREMENT OF PROXIMAL FEMUR STRAIN WITH TOTAL HIP-ARTHROPLASTY [J].
CROWNINSHIELD, RD ;
PEDERSEN, DR ;
BRAND, RA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1980, 102 (03) :230-233
[8]
Fessler H, 1989, Proc Inst Mech Eng H, V203, P1, DOI 10.1243/PIME_PROC_1989_203_003_01
[9]
A TECHNIQUE FOR THE INVITRO MEASUREMENT OF PRINCIPAL STRAINS IN THE HUMAN TIBIA [J].
FINLAY, JB ;
BOURNE, RB ;
MCLEAN, J .
JOURNAL OF BIOMECHANICS, 1982, 15 (10) :723-739
[10]
Femoral surface strain in intact composite femurs: A custom computer analysis of the photoelastic coating technique [J].
Grecula, MJ ;
Morris, RP ;
Laughlin, JC ;
Buford, WL ;
Patterson, RM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (07) :926-933