Infant flow driver or single prong nasal continuous positive airway pressure: short-term physiological effects

被引:31
作者
Ahluwalia, JS [1 ]
White, DK [1 ]
Morley, CJ [1 ]
机构
[1] Addenbrookes Hosp, Neonatal Intens Care Unit, Cambridge CB2 2QQ, England
关键词
CPAP; crossover study; Infant Flow Driver; neonate; oxygenation;
D O I
10.1080/08035259850157408
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The effectiveness of single prong nasal continuous positive airway pressure (CPAP) was compared with the Infant Flow Driver (IFD) in a crossover study in 20 neonates treated with greater than or equal to 30% oxygen by nasal CPAP. They were randomized to the device used at the start of the study. Each infant was studied for four consecutive 2-h periods alternating between single prong nasal CPAP and the IFD. The FiO(2) from the IFD read 0.02 higher than the same setting on the ventilators used for single prong nasal CPAP. The IFD improved the mean (95% CI) of the FiO(2) by 0.05 (0.02-0.08), p = 0.008. Taking into account the systematic error in the FiO(2) between the devices the real mean improvement in FiO(2) produced by the IFD was 0.03 (-0.005 to 0.06), p = 0.09. There were no significant differences in respiratory rate, heart rate, blood pressure or comfort score of infants during periods of single nasal prong CPAP compared with periods on the IFD.
引用
收藏
页码:325 / 327
页数:3
相关论文
共 13 条
[1]   CONTROLLED TRIAL OF CONTINUOUS POSITIVE AIRWAY PRESSURE GIVEN BY FACE MASK FOR HYALINE-MEMBRANE DISEASE [J].
ALLEN, LP ;
REYNOLDS, EOR ;
RIVERS, RPA ;
LESOUEF, PN ;
WIMBERLEY, PD .
ARCHIVES OF DISEASE IN CHILDHOOD, 1977, 52 (05) :373-378
[2]   PROLONGED APNEA OF PREMATURITY - TREATMENT WITH CONTINUOUS AIRWAY DISTENDING PRESSURE DELIVERED BY NASOPHARYNGEAL TUBE [J].
BOROS, SJ ;
REYNOLDS, JW .
CLINICAL PEDIATRICS, 1976, 15 (02) :123-&
[3]  
GOLDMAN SL, 1979, PEDIATRICS, V64, P160
[4]  
GREENOUGH A, 1992, TXB NEONATOLOGY, P385
[5]   TREATMENT OF IDIOPATHIC RESPIRATORY-DISTRESS SYNDROME WITH CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
GREGORY, GA ;
KITTERMAN, JA ;
PHIBBS, RH ;
TOOLEY, WH ;
HAMILTON, WK .
NEW ENGLAND JOURNAL OF MEDICINE, 1971, 284 (24) :1333-+
[6]  
Klausner JF, 1996, PEDIATR PULM, V22, P188, DOI 10.1002/(SICI)1099-0496(199609)22:3<188::AID-PPUL8>3.3.CO
[7]  
2-F
[8]   NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE - EXPERIENCES WITH A NEW TECHNICAL APPROACH [J].
MOA, G ;
NILSSON, K .
ACTA PAEDIATRICA, 1993, 82 (02) :210-211
[9]   A NEW DEVICE FOR ADMINISTRATION OF NASAL CONTINUOUS POSITIVE AIRWAY PRESSURE IN THE NEWBORN - AN EXPERIMENTAL-STUDY [J].
MOA, G ;
NILSSON, K ;
ZETTERSTROM, H ;
JONSSON, LO .
CRITICAL CARE MEDICINE, 1988, 16 (12) :1238-1242
[10]   Nasal deformities resulting from flow driver continuous positive airway pressure [J].
Robertson, NJ ;
McCarthy, LS ;
Hamilton, PA ;
Moss, ALH .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1996, 75 (03) :F209-F212