The determination of phosphorus in a microfluidic manifold demonstrating long-term reagent lifetime and chemical stability utilising a colorimetric method

被引:60
作者
Bowden, A [1 ]
Diamond, D [1 ]
机构
[1] Dublin City Univ, Sch Chem Sci, Natl Ctr Sensor Res, Dublin 9, Ireland
关键词
reagent lifetime; microfluidic manifold; stopped flow; UV-LED; environmental monitoring; phosphate;
D O I
10.1016/S0925-4005(03)00024-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The chemical stability of the vanadomolybdophosphoric acid method for phosphate determination in a microfluidic manifold is described. The reagent lifetime has been shown to extend to more than I year. A stopped flow regime has been implemented, which enabled a very simple microfluidic manifold design to be employed, and has the added advantages of low reagent consumption coupled with less waste generation and access to the complete reaction profile in the optical cuvette on the microfluidic chip. Optical detection was achieved with an UV-LED, integrated into the microfluidic chip holder, coupled to a portable spectrometer via an optical fibre. The manifold includes integration of reagent and sample introduction inlets, a mixing channel and an optical cuvette of 400 mum path length. Two reagent batches were prepared (December 1999 and April 2001) and were shown to still be highly comparable after I year in storage. Multiple calibrations have been performed on the microfluidic system over a 12-month period showing only minimal loss in performance and a standard orthophosphate-containing sample was analysed in the microfluidic manifold on a weekly basis with a relative standard deviation of <2.3%. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:170 / 174
页数:5
相关论文
共 12 条
  • [1] A prototype industrial sensing system for phosphorus based on micro system technology
    Bowden, M
    Sequeira, M
    Krog, JP
    Gravesen, P
    Diamond, D
    [J]. ANALYST, 2002, 127 (01) : 1 - 4
  • [2] Analysis of river water samples utilising a prototype industrial sensing system for phosphorus based on micro-system technology
    Bowden, M
    Sequiera, M
    Krog, JP
    Gravesen, P
    Diamond, D
    [J]. JOURNAL OF ENVIRONMENTAL MONITORING, 2002, 4 (05): : 767 - 771
  • [3] Daridon A, 2000, MESA MG, P303
  • [4] Chemical sensing using an integrated microfluidic system based on the Berthelot reaction
    Daridon, A
    Sequeira, M
    Pennarun-Thomas, G
    Dirac, H
    Krog, JP
    Gravesen, P
    Lichtenberg, J
    Diamond, D
    Verpoorte, E
    de Rooij, NF
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2001, 76 (1-3): : 235 - 243
  • [5] Further studies into the development of a micro-FIA (μFIA) system based on electroosmotic flow for the determination of phosphate as orthophosphate
    Doku, GN
    Haswell, SJ
    [J]. ANALYTICA CHIMICA ACTA, 1999, 382 (1-2) : 1 - 13
  • [6] FREEMANTLE M, 2000, SCIENCE, V77, P27
  • [7] Microfluidics - a review
    Gravesen, Peter
    Branebjerg, Jens
    Jensen, Ole Sondergard
    [J]. Journal of Micromechanics and Microengineering, 1993, 3 (04) : 168 - 182
  • [8] Harrison DJ, 2000, MESA MG, P195
  • [9] MANZ A, 1991, CHIMIA, V45, P103
  • [10] RUZICKA J, 1978, ANAL CHIM ACTA, V99, P37