Development and verification of an automated sample processing protocol for quantitation of human immunodeficiency virus type 1 RNA in plasma

被引:9
作者
Lee, BG
Fiebelkorn, KR
Caliendo, AM
Nolte, FS
机构
[1] Emory Univ Hosp, Clin Labs, Atlanta, GA 30322 USA
[2] Emory Univ, Sch Med, Dept Pathol & Lab Med, Atlanta, GA 30322 USA
关键词
D O I
10.1128/JCM.41.5.2062-2067.2003
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
We developed and verified an automated sample processing protocol for use with the AMPLICOR HIV-1 MONITOR test, version 1.5 (Roche Diagnostics, Indianapolis, Ind.). The automated method uses the MagNA Pure LC instrument and total nucleic acid reagents (Roche Applied Science, Indianapolis, Ind.) to extract human immunodeficiency virus type I (HIV-1) RNA from plasma specimens. We compared the HIV-1 load results for a dilution series (1 to 5 nominal log(10) copies/ml) and 175 clinical specimens processed by the automated method to those for the same samples processed by the manual methods specified by the manufacturer. The sensitivity, dynamic range, and precision of the viral load assay obtained by automated processing of specimens were similar to those obtained by an ultrasensitive manual processing method. The results were highly correlated (R-2, 0.95), and were in close agreement, with a mean difference of 0.09 log(10) (standard deviation, 0.292). The limits of agreement were +/-0.58 log(10) for results for samples processed by both the manual and the automated methods. These performance characteristics were achieved with a smaller sample volume (200 versus 500 mul) and without a high-speed centrifugation step and required only 15 min of labor for a batch of 32 samples. In conclusion, the automated sample preparation protocol can replace both the standard and the ultrasensitive manual methods used with the AMPLICOR HIV-1 MONITOR test and can substantially reduce the labor associated with this test.
引用
收藏
页码:2062 / 2067
页数:6
相关论文
共 7 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] Detection of herpes simplex virus DNA in genital and dermal specimens by LightCycler PCR after extraction using the IsoQuick, MagNA pure, and BioRobot 9604 methods
    Espy, MJ
    Rys, PN
    Wold, AD
    Uhl, JR
    Sloan, LM
    Jenkins, GD
    Ilstrup, DM
    Cockerill, FR
    Patel, R
    Rosenblatt, JE
    Smith, TF
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (06) : 2233 - 2236
  • [3] Fiebelkorn KR, 2002, CLIN CHEM, V48, P1613
  • [4] Kessler HH, 2001, CLIN CHEM, V47, P1124
  • [5] Automated extraction of genomic DNA from medically important yeast species and filamentous fungi by using the MagNA pure LC system
    Loeffler, J
    Schmidt, K
    Hebart, H
    Schumacher, U
    Einsele, H
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (06) : 2240 - 2243
  • [6] HIV viral load markers in clinical practice
    Saag, MS
    Holodniy, M
    Kuritzkes, DR
    OBrien, WA
    Coombs, R
    Poscher, ME
    Jacobsen, DM
    Shaw, GM
    Richman, DD
    Volberding, PA
    [J]. NATURE MEDICINE, 1996, 2 (06) : 625 - 629
  • [7] Evaluation of an automated sample preparation protocol for quantitative detection of hepatitis C virus RNA
    Stelzl, E
    Kormann-Klement, A
    Haas, J
    Daghofer, E
    Santner, BI
    Marth, E
    Kessler, HH
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (04) : 1447 - 1450