Enhanced, simplified glucose sensors: Long-term clinical application of wearable artificial endocrine pancreas

被引:46
作者
Shichiri, M [1 ]
Sakakida, M [1 ]
Nishida, K [1 ]
Shimoda, S [1 ]
机构
[1] Kumamoto Univ, Sch Med, Dept Metab Med, Kumamoto 860, Japan
关键词
wearable artificial endocrine pancreas; miniaturized extracorporeal glucose monitoring system based on microdialysis sampling method; ferrocene-mediated needle-type glucose sensor; poly(2-methacryloyloxyethyl phosphorylcholine-co-n-butyl methacrylate) membrane; subcutaneous insulin infusion algorithm; short acting insulin analogue;
D O I
10.1046/j.1525-1594.1998.06043.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
At present, 2 major problems should be solved before long-term application of the wearable artificial endocrine pancreas, the development of a reliable and stable glucose monitoring system and the development of a subcutaneous insulin infusion algorithm. With either a miniaturized extracorporeal glucose monitoring system based on microdialysis sampling method or a ferrocene-mediated needle-type glucose sensor covered with highly biocompatible membrane, poly(2-methacryloyloxyethyl phosphorylcholine-co-n-butyl methacrylate) (poly[MPC-co-BMA]) membrane, subcutaneous glucose concentrations could be monitored for 7 days without any in vivo calibrations, followed by 14 days with one point calibration. Considering the management and safety of the insulin delivery route, subcutaneous insulin infusion is obligatory. With the subcutaneous insulin infusion algorithm using a short acting insulin analogue (Insulin Lispro), near physiological glycemic control could be established in diabetic patients without showing any delayed hyperinsulinemia or hypoglycemia. The wearable artificial endocrine pancreas is now recognized as an excellent therapeutic tool for regulating blood glucose excursions physiologically in ambulatory diabetic patients on a long-term basis.
引用
收藏
页码:32 / 42
页数:11
相关论文
共 21 条
[1]  
ASAKAWA N, 1987, DIABETES RES CLIN EX, V5, P193
[2]  
GORIYA Y, 1979, MED PROG TECHNOL, V6, P99
[3]   DEVELOPMENT OF A MINIATURIZED GLUCOSE MONITORING-SYSTEM BY COMBINING A NEEDLE-TYPE GLUCOSE SENSOR WITH MICRODIALYSIS SAMPLING METHOD - LONG-TERM SUBCUTANEOUS TISSUE GLUCOSE MONITORING IN AMBULATORY DIABETIC-PATIENTS [J].
HASHIGUCHI, Y ;
UEMURA, T ;
SAKAKIDA, M ;
KAJIWARA, K ;
NISHIDA, K ;
SHICHIRI, M .
DIABETES CARE, 1994, 17 (05) :387-396
[4]   [LYS(B28), PRO(B29)]-HUMAN INSULIN - A RAPIDLY ABSORBED ANALOG OF HUMAN INSULIN [J].
HOWEY, DC ;
BOWSHER, RR ;
BRUNELLE, RL ;
WOODWORTH, JR .
DIABETES, 1994, 43 (03) :396-402
[5]   REDUCED THROMBOGENICITY OF POLYMERS HAVING PHOSPHOLIPID POLAR GROUPS [J].
ISHIHARA, K ;
ARAGAKI, R ;
UEDA, T ;
WATENABE, A ;
NAKABAYASHI, N .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1990, 24 (08) :1069-1077
[6]   HEMOCOMPATIBILITY OF HUMAN WHOLE-BLOOD ON POLYMERS WITH A PHOSPHOLIPID POLAR GROUP AND ITS MECHANISM [J].
ISHIHARA, K ;
OSHIDA, H ;
ENDO, Y ;
UEDA, T ;
WATANABE, A ;
NAKABAYASHI, N .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (12) :1543-1552
[7]   PREPARATION OF PHOSPHOLIPID POLYMERS AND THEIR PROPERTIES AS POLYMER HYDROGEL MEMBRANES [J].
ISHIHARA, K ;
UEDA, T ;
NAKABAYASHI, N .
POLYMER JOURNAL, 1990, 22 (05) :355-360
[8]  
NISHIDA K, 1995, MED PROG TECHNOL, V21, P91
[9]   A MATHEMATICAL INSULIN-SECRETION MODEL AND ITS VALIDATION IN ISOLATED RAT PANCREATIC-ISLETS PERIFUSION [J].
NOMURA, M ;
SHICHIRI, M ;
KAWAMORI, R ;
YAMASAKI, Y ;
IWAMA, N ;
ABE, H .
COMPUTERS AND BIOMEDICAL RESEARCH, 1984, 17 (06) :570-579
[10]   INTENSIVE INSULIN THERAPY PREVENTS THE PROGRESSION OF DIABETIC MICROVASCULAR COMPLICATIONS IN JAPANESE PATIENTS WITH NON-INSULIN-DEPENDENT DIABETES-MELLITUS - A RANDOMIZED PROSPECTIVE 6-YEAR STUDY [J].
OHKUBO, Y ;
KISHIKAWA, H ;
ARAKI, E ;
MIYATA, T ;
ISAMI, S ;
MOTOYOSHI, S ;
KOJIMA, Y ;
FURUYOSHI, N ;
SHICHIRI, M .
DIABETES RESEARCH AND CLINICAL PRACTICE, 1995, 28 (02) :103-117