HPLC determination of Vitamin D3 and its metabolite in human plasma with on-line sample cleanup

被引:33
作者
Brunetto, MR
Obando, MA
Gallignani, M
Alarcón, OM
Nieto, E
Salinas, R
Burguera, JL
Burguera, M
机构
[1] Univ Los Andes, Fac Sci, Venezuelan Andean Inst Chem Res, IVAIQUIM, Merida 5101A, Venezuela
[2] Univ Los Andes, Fac Med, Metab Bone Res Grp, Merida 5101A, Venezuela
关键词
Vitamin D-3; 25-OH-D-3; plasma samples; HPLC; solid-phase extraction; column switching; on-line sample cleanup;
D O I
10.1016/j.talanta.2004.04.035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A HPLC method with automated column switching and UV-diode array detection is described for the simultaneous determination of Vitamin D-3 and 25-hydroxyvitamin D-3 (25-OH-D-3) in a sample of human plasma. The system uses a BioTrap precolumn for the on-line sample cleanup. A sample of 1 ml of human plasma was treated with 2 ml of a mixture of ethanol-acetonitrile (2:1 (v/v)). Following centrifugation, the supernatant was evaporated to dryness under a stream of dry and pure nitrogen. The residue was reconstituted in 250 muL of a solution of methanol 5 mmol l(-1) phosphate buffer, pH 6.5 (4:1 (v/v)), and a 200 mul aliquot of this solution was injected onto the BioTrap precolumn. After washing during 5 min with a mobile phase constituted by a solution of 6% acetonitrile in 5 mmol l(-1) phosphate buffer, pH 6.5 (extraction mobile phase), the retained analytes were then transferred to the analytical column in the backflush mode. The analytical separation was then performed by reverse-phase chromatography in the gradient elution mode with the solvents A and B (Solvent A: acetonitrile-phosphate buffer 5 mmol l(-1), pH 6.5; 20:80 (v/v); solvent B: methanol-acetonitrile-tetrahydrofuran, 65:20:15 (v/v)). The compounds of interest were detected at 265 nm. The method was linear in the range 3.0-32.0 ng ml(-1) with a limit of quantification of 3.0 ng ml(-1). Quantitative recoveries from spiked plasma samples were between 91.0 and 98.0%. In all cases, the coefficient of variation (CV) of the intra-day and inter-day-assay precision was less than or equal to 2.80%. The proposed method permitted the simultaneous determination of Vitamin D-3 and 25-OH-D-3 in 16 min, with an adequate precision and sensitivity. However, the overlap of the sample cleanup step with the analysis increases the sampling frequency to five samples h(-1). The method was successfully applied for the determination of Vitamin D-3 and 25-OH-D-3 in plasma from 46 female volunteers, ranging from 50 to 94 years old. Vitamin D-3 and 25-OH-D-3 concentrations in plasma were found from 4.30-40.70 ng ml(-1) (19.74 +/- 9.48 ng ml(-1)) and 3.1-36.52 ng ml(-1) (7.13 +/- 7.80 ng ml(-1)), respectively. These results were in good agreement with data published by other authors. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:1364 / 1370
页数:7