Enhanced standardization of the International Normalized Ratio through the use of plasma calibrants: a concise review

被引:23
作者
Adcock, DM
Duff, S
机构
[1] Univ Colorado, Hlth Sci Ctr, Colorado Permanente Med Grp, Dept Pathol, Aurora, CO 80011 USA
[2] Precis Biolog, Dartmouth, NS B3B 1P7, Canada
关键词
prothrombin time; International Normalized Ratio; International Sensitivity Index; plasma calibrants;
D O I
10.1097/00001721-200010000-00001
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The INR was introduced in 1983 as a means of standardizing the prothrombin time test and improving management of patients receiving warfarin-based oral anticoagulant therapy. Despite the eventual widespread use of the INR system, unacceptable levels of inter-laboratory variation persist. The use of certified plasma calibrants has been studied as a means to reducing INR variation. This review examines the pre-analytical and analytical variables contributing to INR problems. Also, the findings of several multicenter studies in which plasma calibrants were deployed are presented. Issues such as the number of calibrants used and their composition, calibrant format, plasma citrate concentration, reference thromboplastins used, methods of calibrant certification and alternative techniques for local system calibration are examined. The development of consensus guidelines on the use of plasma calibrants for INR standardization is recommended. Blood Coagul Fibrinolysis 11:583-590 (C) 2000 Lippincott Williams & Wilkins.
引用
收藏
页码:583 / 590
页数:8
相关论文
共 42 条
  • [1] Adcock DM, 1997, AM J CLIN PATHOL, V107, P105
  • [2] ADCOCK DM, UNPUB THROMB HAEMOST
  • [3] Brien WF, 1999, AM J CLIN PATHOL, V111, P193
  • [4] Cai ZJ, 1999, WHO TECH REP SER, V887, P1
  • [5] Calibration of local systems with lyophilized calibrant plasmas improves the interlaboratory variability of the INR in the Italian external quality assessment scheme
    Chantarangkul, V
    Tripodi, A
    Cesana, BM
    Mannucci, PM
    [J]. THROMBOSIS AND HAEMOSTASIS, 1999, 82 (06) : 1621 - 1626
  • [6] ASSESSMENT OF VALUE OF CALIBRATED LYOPHILIZED PLASMAS TO DETERMINE INTERNATIONAL SENSITIVITY INDEX FOR COAGULOMETERS
    CLARKE, K
    TABERNER, DA
    THOMSON, JM
    MORRIS, JA
    POLLER, L
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1992, 45 (01) : 58 - 60
  • [7] *COLL AM PATH, 1994, COMPR PROF TEST SURV
  • [8] THE RELIABILITY OF MANUFACTURER-DETERMINED, INSTRUMENT-SPECIFIC INTERNATIONAL SENSITIVITY INDEX VALUES FOR CALCULATING THE INTERNATIONAL NORMALIZED RATIO
    CUNNINGHAM, MT
    JOHNSON, GF
    PENNELL, BJ
    OLSON, JD
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1994, 102 (01) : 128 - 133
  • [9] Calculation of system international sensitivity index: How many calibrant plasmas are required?
    Dufty, JMK
    Craig, S
    Stevenson, KJ
    Taberner, DA
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1997, 50 (01) : 40 - 44
  • [10] DUNCAN EM, 1994, THROMB HAEMOSTASIS, V72, P84