Absolute calibration of LIDAR Thomson scattering systems by rotational Raman scattering

被引:16
作者
Scannell, R. [1 ]
Beurskens, M. [1 ]
Kempenaars, M. [1 ]
Naylor, G. [1 ]
Walsh, M. [2 ]
O'Gorman, T. [3 ]
Pasqualotto, R. [4 ]
机构
[1] EURATOM CCFE Fus Assoc, Culham Sci Ctr, Abingdon OX14 3DB, Oxon, England
[2] ITER Org, CHD Dept, Div Diagnost, F-13106 Cadarache, St Paul Durance, France
[3] Univ Coll Cork, Dept Elect & Elect Engn, Cork, Ireland
[4] Euratom Enea Fus, Consorzio RFX Assoc, I-35127 Padua, Italy
基金
英国工程与自然科学研究理事会;
关键词
calibration; hydrogen; nitrogen; optical radar; plasma density; plasma diagnostics; plasma toroidal confinement; Raman spectra; Tokamak devices;
D O I
10.1063/1.3374111
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Absolute calibration of LIDAR Thomson scattering systems on large fusion devices may be achieved using rotational Raman scattering. The choice of calibrating gas molecule presents different options and design trade-offs and is likely to be strongly dependent on the laser wavelength selected. Raman scattering of hydrogenic molecules produces a very broad spectrum, however, with far fewer scattered photons than scattering from nitrogen or oxygen at the same gas pressure. Lower laser wavelengths have the advantage that the Raman cross section increases, sigma(Raman)proportional to 1/lambda(4)(0), but the disadvantage that the spectral width of the scattered spectrum decreases, Delta lambda(Raman)proportional to lambda(2)(0). This narrower spectrum makes measurement closer to the laser wavelength necessary. The design of the calibration technique presents a number of challenges. Some of these challenges are generic to all Thomson scattering systems. These include detecting a sufficient number of photoelectrons and designing filters that measure close to the laser wavelength while simultaneously achieving adequate blocking of the laser wavelength. An issue specific to LIDAR systems arises since the collection optics operates over a wide range of depth of field. This wide depth of field has the effect of changing the angle of light incident on the optical interference filter with plasma major radius. The angular distribution then determines the effective spectral transmission function of the interference filter and hence impacts on the accuracy of the absolute calibration. One method that can be used to increase absolute calibration accuracy is collecting both Stokes and anti-Stokes lines with optical filter transmission bands specifically designed to reduce systematic uncertainty.
引用
收藏
页数:8
相关论文
共 15 条
[1]   NONLINEAR-OPTICAL EFFECTS IN RAMAN CALIBRATIONS OF A THOMSON SCATTERING SYSTEM [J].
BASSAN, M ;
GIUDICOTTI, L ;
PASQUALOTTO, R .
APPLIED OPTICS, 1993, 32 (27) :5313-5323
[2]   ITER LIDAR performance analysis [J].
Beurskens, M. N. A. ;
Giudicotti, L. ;
Kempenaars, M. ;
Scannell, R. ;
Walsh, M. J. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10)
[3]  
DESANDE MV, 2002, THESIS TU EINDHOVEN
[4]  
GIUDICOTTI L, NUCL FUSION UNPUB
[5]  
Huber K. P., 1950, MOL SPECTRA MOL STRU
[6]   Enhancement of the JET edge LIDAR Thomson scattering diagnostic with ultrafast detectors [J].
Kempenaars, M. ;
Flanagan, J. C. ;
Giudicotti, L. ;
Walsh, M. J. ;
Beurskens, M. ;
Balboa, I. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10)
[7]   Thomson scattering density calibration by Rayleigh and rotational Raman scattering on NSTX [J].
LeBlanc, B. P. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10)
[8]  
MCCOOL SC, 1981, DOEET530427 U TEX
[9]   ABSOLUTE ROTATIONAL RAMAN CROSS-SECTIONS FOR N2, O2, AND CO2 [J].
PENNEY, CM ;
STPETERS, RL ;
LAPP, M .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1974, 64 (05) :712-716
[10]  
ROUMLHR H, 1981, PHYS LETT A, V81, P451, DOI DOI 10.1016/0375-9601(81)90409-6