Secondary emission monitor for low-interception monitoring (SLIM): An innovative nondestructive beam monitor for the extraction lines of a hadrontherapy center

被引:6
作者
Badano, L [1 ]
Ferrando, O
Klatka, T
Koziel, M
Molinari, G
Pezzetta, M
机构
[1] TERA Fdn, Milan, Italy
[2] CERN, Geneva, Switzerland
[3] AGH Univ Sci & Technol, Dept Elect, Krakow, Poland
关键词
D O I
10.1109/TNS.2004.839101
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Real-time monitoring of hadrontherapy beam intensities and profiles is a critical issue for the optimization of the dose delivery to the patient tumour tissue, the patient safety, and the operation of the accelerator complex. For this purpose, an innovative beam monitor, based on the secondary emission of electrons by a nonperturbative, submicrometer-thick Al target placed directly in the extracted beam path, is being proposed. The secondary electrons, accelerated by an electrostatic focusing system, are detected by a monolithic silicon position sensitive sensor, which provides the beam intensity and position with a precision of 1 mm at a 10-kHz frame rate. The conceptual design and the engineering study optimized for hadrontherapy are presented together with the results of the laboratory tests with thermoionic electrons and the first measurements in a hadron beam.
引用
收藏
页码:2990 / 2998
页数:9
相关论文
共 16 条
[1]  
AMALDI U, 1994, P 1 INT S HADR COM
[2]  
BADANO L, SECONDARY EMISSION S
[3]  
BADANO L, 2003, P 6 EUR WORKSH BEAM, P77
[5]  
BOSSER K, 1987, 8706 SPSABM CERN
[6]   Development, fabrication and test of a highly segmented hybrid photodiode [J].
Braem, A ;
Chesi, E ;
Joram, C ;
Séguinot, J ;
Weilhammer, P ;
Ypsilantis, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 478 (1-2) :400-403
[7]   Silicon Ultra fast Cameras for electron and γ sources In Medical Applications [J].
Caccia, M ;
Airoldi, A ;
Alemi, M ;
Amati, M ;
Badano, L ;
Bartsch, V ;
Berst, D ;
Bianchi, C ;
Bol, H ;
Bulgheroni, A ;
Cannillo, F ;
Cappellini, C ;
Czermak, A ;
Claus, G ;
Colledani, C ;
Conte, L ;
Deptuch, G ;
de Boer, W ;
Dierlamm, A ;
Domanski, K ;
Dulinski, W ;
Dulny, B ;
Ferrando, O ;
Grigoriev, E ;
Grabiec, P ;
Lorusso, R ;
Jaroszewicz, B ;
Jungermann, L ;
Kucewicz, W ;
Kucharski, K ;
Kuta, S ;
Leo, G ;
Machowski, W ;
Marczewski, J ;
Mondry, G ;
Niemiec, H ;
Novario, R ;
Paolucci, L ;
Pezzetta, M ;
Popowski, Y ;
Riester, JL ;
Sampietro, C ;
Sapor, M ;
Schweickert, H ;
Spanò, B ;
Tomaszewski, D ;
Zalewska, A .
NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 2003, 125 :133-138
[8]   WEAKLY IONIZING CHARGED-PARTICLE DETECTORS WITH HIGH-EFFICIENCY USING TRANSITORY ELECTRONIC SECONDARY-EMISSION OF POROUS CSI [J].
CHIANELLI, C ;
AGERON, P ;
BOUVET, JP ;
KAROLAK, M ;
MARTIN, S ;
ROBERT, JP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 273 (01) :245-256
[9]  
DAHL DA, IDAHO NAT ENG ENV LA
[10]   Design and testing of Monolithic Active Pixel Sensors for charged particle tracking [J].
Deptuch, G ;
Berst, JD ;
Claus, G ;
Colledani, C ;
Dulinski, W ;
Gornushkin, Y ;
Husson, D ;
Riester, JL ;
Winter, M .