Fully Wireless Implantable Cardiovascular Pressure Monitor Integrated with a Medical Stent

被引:161
作者
Chow, Eric Y. [1 ]
Chlebowski, Arthur L. [1 ]
Chakraborty, Sudipto [3 ]
Chappell, William J. [2 ]
Irazoqui, Pedro P. [1 ]
机构
[1] Purdue Univ, Brain Comp Interface Lab, Weldon Sch Biomed Engn, W Lafayette, IN 47907 USA
[2] Purdue Univ, Sch Elect & Comp Engn, IDEAS Microwave Lab, W Lafayette, IN 47907 USA
[3] Texas Instruments Inc, Commun Syst Lab, Dallas, TX 47907 USA
关键词
Biomedical applications of electromagnetic radiation; biomedical monitoring; biomedical telemetry; implantable biomedical devices; HEART-FAILURE; DIELECTRIC-PROPERTIES; HEMODYNAMIC MONITOR; BIOLOGICAL TISSUES; MULTICENTER; EXPERIENCE; FREQUENCY; SURVIVAL; DESIGN;
D O I
10.1109/TBME.2010.2041058
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a fully wireless cardiac pressure sensing system. Food and Drug Administration (FDA) approved medical stents are explored as radiating structures to support simultaneous transcutaneous wireless telemetry and powering. An application-specific integrated circuit (ASIC), designed and fabricated using the Texas Instruments 130-nm CMOS process, enables wireless telemetry, remote powering, voltage regulation, and processing of pressure measurements from a microelectromechanical systems (MEMS) capacitive sensor. This paper demonstrates fully wireless-pressure-sensing functionality with an external 35-dB.m RF powering source across a distance of 10 cm. Measurements in a regulated pressure chamber demonstrate the ability of the cardiac system to achieve pressure resolutions of 0.5 mmHg over a range of 0-50 mmHg using a channel data-rate of 42.2 kb/s.
引用
收藏
页码:1487 / 1496
页数:10
相关论文
共 64 条
[1]  
Ahlbom A, 1998, HEALTH PHYS, V74, P494
[2]  
Allen M., 2005, Implantable Wireless Sensor for Blood Pressure Measurement with an Artery
[3]  
[Anonymous], 2004, HEART DIS STROKE STA
[4]  
[Anonymous], 1999, IEEE STANDARD C951
[5]  
[Anonymous], 802 IEEE
[6]  
BALANIS CA, 1989, ADV ENG ELECTROMAGNE, P104
[7]   Development of implantable devices for continuous ambulatory monitoring of central hemodynamic values in heart failure patients [J].
Bennett, T ;
Kjellstrom, B ;
Taepke, R ;
Ryden, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (06) :573-584
[8]  
Bensky A., 2000, Short-Range Wireless Communication, V1st, P324
[9]   THE HISTORY OF POWER TRANSMISSION BY RADIO-WAVES [J].
BROWN, WC .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1984, 32 (09) :1230-1242
[10]  
Chen L.J., 2004, SILICIDE TECHNOLOGY