Experimental setup for the determination of analytes contained in ultrasonically levitated drops

被引:20
作者
Rohling, O [1 ]
Weitkamp, C [1 ]
Neidhart, B [1 ]
机构
[1] GKSS Forschungszentrum Geesthacht GmbH, Inst Phys & Chem Analyt, D-21502 Geesthacht, Germany
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 2000年 / 368卷 / 2-3期
关键词
D O I
10.1007/s002160000391
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated technique for the determination of analytes in an ultrasonically levitated sample of 2 to 5 mu L volume has been developed. Contactless dosing of reagents or solvents into an ultrasonically levitated drop was realized via piezoelectric micropumps. Drop size was continually controlled with a programmed CCD camera. A diode array spectrometer designed for the use with optical fibers was used for absorption and fluorescence measurements. Determinations via direct absorption measurements following the method of standard addition and acid-base titrations with an absorption indicator and a fluorescent indicator were carried out. The mean consumption of sodium hydroxide added via a piezoelectric micropump in five successive titrations with 18 nmol of sulfuric acid inside a levitated drop (indicator bromothymol blue) had a relative standard deviation of 0.7 % and differed only by 0.2 % from the expected value.
引用
收藏
页码:125 / 129
页数:5
相关论文
共 7 条
[1]   Acoustic levitation device for sample pretreatment in microanalysis and trace analysis [J].
Eberhardt, R ;
Neidhart, B .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 365 (06) :475-479
[2]  
EBERHARDT R, ULTRASCHALL LEVITATI
[3]  
Ortner HM, 1996, FRESEN J ANAL CHEM, V355, P657
[4]  
SKOUTERIS D, 1996, THESIS PHILIPPS U MA
[5]   Evaluation of a mirror-polishing technique for fluorocarbon polymer surfaces for reduction of contamination from containers used in ultratrace analysis [J].
Takenaka, M ;
Hayashi, M ;
Suzuki, I ;
Yamada, Y ;
Takamatsu, K ;
Kageyama, M .
ANALYTICAL CHEMISTRY, 1997, 69 (05) :972-976
[6]   Acoustically levitated droplets - A new tool for micro and trace analysis [J].
Welter, E ;
Neidhart, B .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 357 (03) :345-350
[7]   OPTICAL FIBER TITRATIONS .3. CONSTRUCTION AND PERFORMANCE OF A FLUOROMETRIC ACID - BASE TITRATOR WITH A BLUE LED AS A LIGHT-SOURCE [J].
WOLFBEIS, OS ;
SCHAFFAR, BPH ;
KASCHNITZ, E .
ANALYST, 1986, 111 (11) :1331-1334