Metrics of resolution and performance for CD-SEMs

被引:58
作者
Joy, DC [1 ]
Ko, YU [1 ]
Hwu, JJ [1 ]
机构
[1] Univ Tennessee, EM Facil, Knoxville, TN 37996 USA
来源
METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XIV | 2000年 / 3998卷
关键词
scanning electron microscope; resolution; signal to noise; probe size; Fourier transform;
D O I
10.1117/12.386444
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The performance of scanning electron beam instruments such as CD-SEMs can be defined in terms of parameters such as the beam probe size, the spatial resolution, and the signal to noise ratio of the image. A knowledge of these quantities is important in verifying the fact that an instrument meets its specification, and subsequently for tracking and optimizing its performance during use. Analytical methods based on the power spectrum (2-D Fourier transform analysis) of images are now beginning to be used for these purposes but care must be exercised to ensure reliable and meaningful results. Two new methods are suggested which can offer more detailed information about the microscope performance while avoiding the pitfalls of the simpler technique. Code implementing these tests, written as a plug-in macro for the well known NIH Image program, is available on-line.
引用
收藏
页码:108 / 114
页数:7
相关论文
共 10 条
[1]  
DODSON TA, 1990, P 12 INT C EM, P406
[2]  
ERASMUS SJ, 1980, I PHYS C SER, V52, P73
[3]  
FRANK J, 1970, BERICH BUNSEN GESELL, V74, P1105
[4]  
Frank J, 1980, COMPUTER PROCESSING, P187
[5]  
GOLDSTEIN JI, 1992, PRACTICAL SCANNING E
[6]  
Joy DC, 1996, SCANNING, V18, P533, DOI 10.1002/sca.4950180802
[7]  
POSTEK M, 1999, COMMUNICATION
[8]  
*SEM IND ASS, 1999 INT TECHN ROADM
[9]   Image sharpness measurement in scanning electron microscopy - Part II [J].
Vladar, AE ;
Postek, MT ;
Davidson, MP .
SCANNING, 1998, 20 (01) :24-34
[10]  
YOUN HK, 1997, P MICR SOC CAN, V24, P63