Validation of home blood glucose meters with respect to clinical and analytical approaches

被引:86
作者
Brunner, GA
Ellmerer, M
Sendlhofer, G
Wutte, A
Trajanoski, Z
Schaupp, L
Quehenberger, F
Wach, P
Krejs, GJ
Pieber, TR
机构
[1] Karl Franzens Univ Graz, Dept Internal Med, Div Diabet & Metab, A-8036 Graz, Austria
[2] Karl Franzens Univ Graz, Dept Med Informat Stat & Documentat, A-8036 Graz, Austria
[3] Graz Univ Technol, Dept Biophys, A-8010 Graz, Austria
[4] Graz Univ Technol, Inst Biomed Engn, A-8010 Graz, Austria
[5] Ludwig Boltzmann Inst Tech Lebenshilfen, Graz, Austria
关键词
D O I
10.2337/diacare.21.4.585
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE- To evaluate the clinical and analytical accuracy of home blood glucose meters. RESEARCH DESIGN AND METHODS- Six blood glucose meters-Reflolux S (Boehringer Mannheim, Mannheim, Germany), One Touch II (LifeScan, Milpitas, CA), Glucocard Memory (Menarini, Florence, Italy), Precision QID (Medisense, Cambridge, U.K.), HaemoCue (HaemoCue, Angelholm, Sweden), and Accutrend alpha (Boehringer Mannheim, Mannheim, Germany)-were compared with a reference method (Beckman Glucose Analyzer II) under controlled conditions (glucose clamp technique). Validation of the blood glucose meters was accomplished by clinically oriented approaches (error grid analysis), statistical approaches (variance components analysis), and by the criteria of the American Diabetes Association (ADA), which recommend a target variability of <5%. RESULTS- A total of 1,794 blood glucose monitor readings and 299 reference values ranging from 2.2 to 18.2 mmol/l were analyzed (705 readings <3.89 mmol/l, 839 readings between 3.89 and 9.99 mmol/l, and 250 readings >9.99 mmol/l). According to error grid analysis, only Reflolux S and Glucocard M had 100% of estimations within the clinically acceptable zones A and B. Assessment of analytical accuracy revealed substantial differences between the glucose meters after separation of the data into defined glycemic ranges. None of the devices met the ADA criteria. CONCLUSIONS- To evaluate accuracy of blood glucose meters, error grid analysis, as well as statistical models, are helpful means and should be performed together. Analytical performance of currently available home blood glucose meters differs substantially within defined glycemic ranges.
引用
收藏
页码:585 / 590
页数:6
相关论文
共 20 条
  • [1] [Anonymous], 1996, DIABETES CARE, V19, pS62
  • [2] ASHWORTH L, 1992, CLIN CHEM, V38, P1479
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    [J]. DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [5] Understanding error grid analysis
    Cox, DJ
    GonderFrederick, LA
    Kovatchev, BP
    Julian, DM
    Clarke, WL
    [J]. DIABETES CARE, 1997, 20 (06) : 911 - 912
  • [6] COMPARISON OF 5 GLUCOSE METERS FOR SELF-MONITORING OF BLOOD-GLUCOSE BY DIABETIC-PATIENTS
    GIFFORDJORGENSEN, RA
    BORCHERT, J
    HASSANEIN, R
    TILZER, L
    EAKS, GA
    MOORE, WV
    [J]. DIABETES CARE, 1986, 9 (01) : 70 - 76
  • [7] Reservations on the use of error grid analysis for the validation of blood glucose assays
    Gough, DA
    Botvinick, EL
    [J]. DIABETES CARE, 1997, 20 (06) : 1034 - 1036
  • [8] INTENSIVE INSULIN THERAPY FOR TREATMENT OF TYPE-I DIABETES
    HIRSCH, IB
    FARKASHIRSCH, R
    SKYLER, JS
    [J]. DIABETES CARE, 1990, 13 (12) : 1265 - 1283
  • [9] HOCKING RR, 1984, ANAL LINEAR MODELS, P316
  • [10] ARTERIAL, ARTERIALIZED VENOUS, VENOUS AND CAPILLARY BLOOD-GLUCOSE MEASUREMENTS IN NORMAL MAN DURING HYPERINSULINEMIC EUGLYCEMIA AND HYPOGLYCEMIA
    LIU, D
    MOBERG, E
    KOLLIND, M
    LINS, PE
    ADAMSON, U
    MACDONALD, IA
    [J]. DIABETOLOGIA, 1992, 35 (03) : 287 - 290